Question 1

Question
Run-outs from horizontal mains or branches shall be taken off ___ the centerline of the pipeline
Answer
  • Above
  • Below
  • At
  • Any of the above

Question 2

Question
While brazing medical gas copper joints, a purge gas shall flow continuously to prevent the formation of copper oxide on the inside surface of the joint. What shall be used as a purge gas?
Answer
  • Oil Free Dry Argon MD
  • Oil Free Dry Helium NF
  • Oil Free Dry Nitrogen NF
  • Any Inert Gas

Question 3

Question
Which of the following filler metals shall be used to braze copper to copper joints in medical gas/vacuum distribution piping?
Answer
  • BAg Series
  • BAu Series
  • BCup Series
  • BNi Series

Question 4

Question
Tubing for positive pressure medical gas systems delivered from the manufacturer shall be _____
Answer
  • Plugged or capped
  • Harddrawn seamless
  • Identified as "oxy" or "med"
  • All of the above

Question 5

Question
In positive pressure systems area alarm sensors for vital life support and critical care areas shall be located on which side of the zone valve box assemblies.
Answer
  • Patient side
  • Outlet/Inlets side
  • Downstream side
  • All of the above

Question 6

Question
The master alarm systems shall activate when the pressure in the main supply lines increase or decrease by _____ percent.
Answer
  • 10
  • 20%
  • 40
  • 50

Question 7

Question
All medical gas alarm system shall have a _______ audible indication for each conditioned monitor
Answer
  • Adjustable
  • Cancelable
  • Optionable
  • All of the above

Question 8

Question
The master alarm systems shall consists of two or more alarm panels, a centralized computer system shall not be permitted to be used as a substitute for one of the required master alarm panels.
Answer
  • True
  • False

Question 9

Question
Since dew point and carbon monoxide monitors are located upstream of the source valve, demand check fittings are not required.
Answer
  • True
  • False

Question 10

Question
Where shall pressure indicators be provided for medical gas systems.
Answer
  • On the main supply
  • At the area alarm
  • Outlet/Inlets side of the zone valve
  • All of the above

Question 11

Question
The scale range of positive pressure analog indicators shall have the normal reading fall within the ___ percentage of the gauge scale
Answer
  • 25
  • 50
  • 75
  • 100

Question 12

Question
Station inlets for WAGD service shall be legibly marked to identify them and shall be of a type appropriate for the flow and vacuum level required by the facility gas anesthetic machine.
Answer
  • True
  • False

Question 13

Question
The factory installed tubing on station outlets/inlets shall extend no further than ____ from the body of the terminal
Answer
  • 6
  • 8
  • 10
  • 12

Question 14

Question
Each station outlet/inlet for medical gas or vacuum shall be ____
Answer
  • Interchangeable
  • Gas-specific
  • Threaded
  • Quick-coupler

Question 15

Question
What is name given to the shutoff valve installed on all lateral branch piping form a main or riser?
Answer
  • Riser
  • Service
  • Zone

Question 16

Question
A shutoff valve shall be placed at the immediate connection of each source system to the distribution piping system. What is the name given to the valve?
Answer
  • Main
  • Supply
  • Source
  • Isolation

Question 17

Question
All valves except zone valve box assemblies shall be
Answer
  • In a secured areas
  • locked or latched
  • Labeled appropriately
  • All of the above

Question 18

Question
Where instrument air system are provided with standby headers
Answer
  • True
  • False

Question 19

Question
What is the minimum number of compressor required in an instrument air system that employs a standby headers?
Answer
  • 0
  • 1
  • 2
  • 3

Question 20

Question
Instrument air system shall be permitted to interconnect with medical air systems that have at least two or more compressors.
Answer
  • True
  • False

Question 21

Question
The quality of the instruments air shall be ___.
Answer
  • Filtered to .01 micron
  • Free of liquids
  • Free of hydrocarbons vapors
  • All of the above

Question 22

Question
A level 1 dedicated WAGD source shall consist of ____or more producers, each shall be sufficient to serve the peak calculated demand.
Answer
  • one
  • Two
  • Three
  • All of the above

Question 23

Question
Vacuum receivers shall be capable of withstanding a gauge vacuum of ____ inches of mercury.
Answer
  • 29
  • 29.9
  • 30
  • All of the above

Question 24

Question
Level 1 medical-surgical vacuum sources shall consist of an automatic means to prevent ____ through any of-cycle vacuum pump.
Answer
  • Pressure loss
  • Back pressure
  • Backflow
  • All of the above

Question 25

Question
Carbon monoxide shall activate the required alarms when the CO level exceeds ____.
Answer
  • 5 ppm
  • 10 ppm
  • 15 ppm
  • 20 ppm

Question 26

Question
Where charcoal filters are required, a color-metric hydrocarbon indicator shall be provided.
Answer
  • True
  • False

Question 27

Question
Where liquid ring compressors are used, air receivers shall be equipped with a high water level sensor that activates a ____alarm indicator.
Answer
  • Master
  • Area
  • Local
  • All of the above

Question 28

Question
The compressor air intakes shall be located outdoors above roof level at a minimum distance of ____ above ground.
Answer
  • 40 ft
  • 30 ft
  • 20 ft
  • 10 ft

Question 29

Question
Where medical air piping systems at different operating air pressures are required, the piping shall separate after the _____.
Answer
  • Filters
  • Final regulators
  • Riser valves
  • Source Valve

Question 30

Question
If dryers, filters and regulators are sized to serve the peak calculated demand, they shall not be duplexed.
Answer
  • True
  • False

Question 31

Question
Level 1 Medical air compressors shall be sized to served the peak calculated demand and in no case shall there be fewer than ____ compressors.
Answer
  • One
  • Two
  • Three
  • Four

Question 32

Question
Receivers for medical air shall be equipped with ____.
Answer
  • Relief valve
  • Automatic and manual drains
  • Sight glass
  • All of the above

Question 33

Question
______ connectors shall connect the air compressors with their intake and outlet piping.
Answer
  • Anti-vibration
  • Dielectric
  • Insulated
  • Flexible

Question 34

Question
Compressors for medical air shall be designed to prevent introduction of contaminants or liquid into the piping systems.
Answer
  • True
  • False

Question 35

Question
Medical air sources shall be used only for ____and calibration of medical respiratory devices.
Answer
  • Human respiration
  • Air-operating device
  • Instrument air
  • All of the above

Question 36

Question
Where the bulk cryogenic liquid oxygen supply systems is outside of and remote from the building served a ___connection shall be installed.
Answer
  • Maintenance cleanout
  • Refilling
  • Emergency oxygen supply
  • All of the above

Question 37

Question
An in building emergency reserve system can be substituted for the EOSC.
Answer
  • True
  • False

Question 38

Question
Bulk cryogenic liquid system site shall have allowance for at least ___ of clearance around storage container, vaporizer and the pressure regulating manifold.
Answer
  • 2 ft
  • 3 ft
  • 4 ft
  • 5 ft

Question 39

Question
Installers of bulk cryogenic liquid systems shall be qualified to meet CGA M-1 Standard, Guide for Medical Gas Installations at Consumer Sites.
Answer
  • True
  • False

Question 40

Question
manifolds for cryogenic liquid cylinder shall indicate when or at a predetermined set point before the reserve contents fall to one day's average supply, indicating ____.
Answer
  • Reserve is in use
  • Primary empty
  • Secondary empty
  • Reserve low

Question 41

Question
Manifolds for cryogenic liquid cylinders shall consist of ____ equal headers, each having sufficient number of liquid containers for an average day's supply.
Answer
  • 4
  • 3
  • 2
  • 1

Question 42

Question
Manifolds for gas cylinders without reserve shall consist of ____ equal headers, each with an average days supply, but not fewer than ____ connections.
Answer
  • Four-four
  • Three-three
  • Two-two
  • One-one

Question 43

Question
Relief valves shall be vented to the ____, except compressed air systems having less than 3000 cu ft at STP
Answer
  • Floor
  • Sewer Drain
  • Back into the systems
  • Outside

Question 44

Question
All pressure relief valve shall be made of brass, bronze or ___.
Answer
  • Stainless
  • Carbon steel
  • Cast Iron
  • Copper

Question 45

Question
Louvered openings for natural ventilation shall have a minimum free area of ___.
Answer
  • 60 sq in
  • 62 sq in
  • 70 sq in
  • 72 sq in

Question 46

Question
Locations for central supply systems and the storage of medical gases shall be secured with lockable doors or gates.
Answer
  • True
  • False

Question 47

Question
Cylinders in use and in storage shall be prevented from reaching temperatures in excess of ___.
Answer
  • 100 F
  • 130 F
  • 160 F
  • 190 F

Question 48

Question
Which of the following systems shall not be permitted to be located in the same indoor enclosure?
Answer
  • Bulk cryogenic liquid system
  • Manifolds for cryogenic liquid cylinders
  • Manifolds for gas cylinders with or without reserve
  • instrument air standby leader

Question 49

Question
liquid containers shall have additional product identification visible from all directions with a minimum of ____inch high letters such as 360 degree wrap around tape for medical liquid containers
Answer
  • 1 in
  • 2 in
  • 3 in
  • 4 in

Question 50

Question
Which of the following hazards shall be considered in the design, installation, testing and maintenance of medical gas and vacuum piping systems?
Answer
  • Contaminations
  • Fire and Smoke
  • Fire and explosions
  • Water and Gas
Show full summary Hide full summary

Similar

Reducing the Impact of Earthquakes
siobhan.quirk
Geometry Formulas
Selam H
Biology AQA 3.1.2 Proteins
evie.daines
Chemistry
Holly Bamford
Main People in Medicine Through Time
Holly Bamford
B3, C3, P3
George Moores
Chemistry (C3)
Amy Lashkari
Health and Social Care
Kelsey Phillips
untitled 2
lola_smily
Of Mice and Men Plot Overview
Landon Valencia
The Gastrointestinal System- Physiology- PMU
Med Student