Computed Radiography & Digital Radiography
Folien in diesem Satz (17)
-
- Image
- Receptor:
- –Film
- screen
- –Computed
- Radiography
- –Direct
- Digital Radiography (DDR/DR)
- The
- computer processing means that systems have a far greater latitude compared to
- film/screen. But
- knowing what happens when you set/manipulate exposure factors still matters
- Increasing kVp also
- increases beam quantity to a
- certain extent: because the electrons with more energy are more efficient at
- removing K-shell electrons. Intensity
- ∝ kVp2
- CR has a tube potential (kVp) that determines the subject contrast in the resultant image - this is connected to K- edge of the receptors
- Anatomical tissues vary in absorption and
- transmission create a range of dark and light areas
- Remember it is the degree of
- photoelectric absorption that determines the range of greys visible on the
- radiographic image. This degree of absorption is strongly dependent on both the
- density and inversely proportional to beam energy
- This is what creates the radiographic
- image. Since it is the interaction of the outer electrons that usually
- influence the degree of absorption, transmission, or scatter, the radiographic
- can be thought of as a map of electron density
- And whilst CR and DR has a wide exposure latitude if there isn't enough information to make a diagnostic image then there isn't enough!
-
- Formally
- it is the tonal range generated within an exposure.
- Sometimes
- called a characteristic curve, that has axes of transmitted light (ie
- range of greys very light to very dark), and exposure (logarithmic
- Sometimes
- called the dynamic range
- In
- digital imaging, changing kVp affects the amount of
- radiation reaching the image receptor and
- affects subject
- contrast (differential attenuation or inherent contrast)
- In
- digital imaging, brightness and image contrast are primary
- controlled during computer processing – but you still need to be aware of the
- effect of kVp on
- subject (inherent) contrast

- Bildunterschrift: Speed of ‘film’ is related to the gradient – a faster film has a steeper gradient, historically because the crystals which made up the film were bigger, requiring less exposure to oxidise to silver. A slower film therefore had smaller crystals requiring more exposure per unit time to generate the same degree of oxidation
-
- Film screen vs CR
- CR
- and DR are digital imaging modalities. Both CR and DR are relative newcomers to
- x-ray imaging
- Film-screen
- uses a cassette similar to the cassette used in CR
- Film-screen
- uses a processor with chemicals where as CR uses a digital processor
- In
- CR the film and intensifying screen is replaced with a phosphor plate
- CR equipment:
- Cassette
- and phosphor plate, Laser
- Image Reader, PACS
- CR Advantages
- –A
- digital image is generated
- –Ability
- to retrofit to existing radiography equipment
- –Mobile
- radiography is easily accomplished
- –Excellent
- image quality
- –Initially
- less expensive than DR
- CR disadvantages
- –Still
- have to use an imaging plate
- –No
- real time saving benefit over traditional radiography
- –Need
- to purchase an imaging reader
- Imaging Plate
- Average
- life cycle over 10000 uses
- Two
- year warranty
- Average
- cost £ 1000 each
- Flexible
- Durable
- High
- Absorption efficiency - high
- % of incoming x-rays effectively contribute to information capture, therefore
- enabling x-ray dose reduction
- Smooth
- Shiny surface
- Correct
- & Incorrect sides? TUBE side
- 35 x 43 cm, 35 x 35 cm, 24 x 30 cm, 18 x 24 cm, 15 x 30 cm
- Storage: Upright ( NOT flat) to prevent pressure on the Phosphor plate. Must not be left unprotected in the x-ray room, sensitive to scatter.
- Cleaning: Every TWO weeks and erased. If not there is an increased risk of becoming stuck in the digitizer.
- Cassette Erasure –Erased ONCE a week at least, UHL it is done first thing every day. Reduce risk of ghost image.
-
- Phosphor plate: Photo-stimulable storage phosphor plate, that stores the
- latent image, is a re-usable detector.
- Two
- part process:
- 1.converting incoming x-rays to light
- photons - image capture
- 2.converting light into electrical signal - image readout
- Laser Image Reader: Scanning mechanism to
- extract the latent image from the plate.
- Transmitted to network for
- review/reporting etc
- Through the PACS system
- Phosphor
- plate is analogous to film – it is sensitive to x-ray irradiation, whereas film
- is not; BUT to demonstrate what happens during an x-ray exposure instead of an intensifying screen – a scintilator
- which converts x-ray to visible light which exposes a film, the light is
- trapped within the plate to be read out later

-

- Protective layer - thin
- & transparent, tough
- and clear. Protects the phosphor layer during handling
- from damage.
- Phosphor layer - active
- component, barium fluorohalide
- crystals doped with europium, contains
- colour centres or F centres, which trap electrons during an exposure.
- Support
- layer - semi
- rigid, give
- strength to the plate
- Conductor
- layer - absorbs
- and removes static electricity
- Light
- shield layer - reflective
- layer, sends
- light in forward direction when released in the reader, ensuring as much light
- as possible is recorded
- Backing
- layer - protects the back of the cassette.
-
- •Exit
- photons are absorbed by the atoms of the phosphor and europium
- •This
- is by the photoelectric effect
- Phosphor plate - image capture
- The absorbed energy excites the electrons
- They are raised to a higher energy state
- Here they become stored/trapped in colour (F) centres
- This is what forms the latent image
- The number of electrons trapped in the F centres is directly related to X-ray beam intensity
- This energy ( latent image ) can be stored for several hours
- However the image will lose about 25%of its energy in 7-8 hours
- The latent image is processed by loading the cassette into an image reader.
- Image reader unit (digitiser)
- The cassette is placed into the reader
- unit
- The imaging plate is extracted and
- scanned
- The image plate is scanned with a finely
- focussed helium-neon laser
- beam.
- The laser releases the trapped electrons
- and they return to a lower energy state.
- This releases energy as visible light.
-

- The
- varying electrical signals are analogue, converted by an ADC to a digital image.
- Once
- the plate is read it is erased to remove all trace of the latent image
- By
- exposing the plate with intense white light to release any trapped electrons.
- Imaging
- plates are sensitive to `Scatter radiation` and should be erased once a week to
- prevent background signal build up. The readers have an erase mode
- The scanning of the plate results in a
- continuous pattern of light intensities sent to a photomultiplier tube (which
- is a photo detector)
- Photomultiplier tube
- collects, amplifies and coverts the light to an electrical signal
- The electrical signal is proportional to
- the range of energies stored in the image plate.
-
- A
- digital image is made up of a 2D array of numbers called a Matrix
- Bit depth - A
- descriptor of the grey scale value attributed to each picture element
- Pixels are the smallest component of a matrix
- The density of each Pixel ( picture
- element) has to be a whole number. Each pixel has a bit depth or a number of
- bits which controls the brightness of the pixel – 4bit, 8 bit, 10 bit or 12 bit
- The greater the bit depth the more the
- precise digitalization of the analogue signal, and the greater the various
- shades of grey
- This is turn improves its contrast
- resolution - a previous slide suggested
- that 1024 discrete shades of grey can be displayed - but we should consider actually how shades of
- grey can the eye distinguish under optimum image display conditions?
- It
- is now established that the human eye can discriminate 'between 700 and 900
- simultaneous shades of gray for
- the available luminance range of current medical displays and in optimal
- conditions.' It is presently felt that it is of 'no use to simultaneously
- display more than 10 bits of gray (1,024 gray shades) because this already exceeds the
- capabilities of the human visual system
- Larger the matrix size, greater the
- number of pixels, better the image quality
-
- Now we have an image, what happens next?
- PACS – Picture Archiving and
- Communication System
- Images are captured by a “ Image Server
- ”. PACS does not generate any images, simply handles the data.
- The Server is the primary point of entry. It
- receives digital signals from all radiology modalities, CT, MR, NM, C-arm, DSA,
- digitisers…
- The language of medical digital imaging is
- called “DICOM”.
- DICOM allows various components to communicate
- with each other.
- In PACS these images are captured as data
- frames and routed through a server to digital display monitor.
- For more information please see resources
- on blackboard
-
- CR imaging screens are ‘doped’ usually
- Barium Fluorohalide;
- with a small quantity of Europium.
- The addition of Eu
- causes defects within the crystal lattice which allows electrons to become
- trapped more efficiently (F-Centre)
- Incident x-ray is absorbed by the BaFH:Eu
- phosphor – which oxidises the Eu atoms, the electrons liberated by this
- process become trapped in a higher energy metastable
- state
- This is the ‘latent’
- image
- The number of trapped electrons is
- proportional to the intensity of the incident x-ray beam
- Flat panel: Caesium Iodide or Selenium
- detector ( FPD)
- Absorbs x-rays directly and converts into
- a digital signal in 4 - 8 seconds
- Indirect flat panel detectors use an intensifying screen or scintillator to convert x-rays to visible light, and then to a photodetector to produce a visible image
- Direct flat panel detectors use incident
- x-rays to generate free electrons which generate the ‘latent’ image directly.
- Selenium is commonly used as a
- photoconductor
- A large voltage difference is applied
- across the array at the same time, which causes the electrons to move to their
- detectors.
-

- Caesium Iodide doped detectors are used as indirect flat panel detectors – there is a intermediate step in the generation of a digital signal.
- Incident x-ray photons interact via photoelectric absorption
- This leads to scintillation
- The light is detected by a photodiode
- The intensity of the light is directly
- proportional to the number of incident x-ray photons
-
Blurring and spatial resolution

- a)Internal
- reflection and diffusion processes
- b) the electron can wander about ionising
- as it goes and losing energy, or
- c)Can be absorbed by another atom leading
- to another ionisation type event, with corresponding re-emission of light. This
- would simulate x-ray incidence additional to the primary point
- d)How can you tell which is which – You
- cant, since the number of interactions is a phenomenal amount (259 * 6.023
- x10exp23) ions present in 1 mole of CsI
- e)Approx ten times the number of stars in the known
- universe (1 x 10 exp 24)

- Bildunterschrift: d) Wide range of intensity e) Narrower range, but decreased amplitude –so less light available f) So structured phosphor is the way to go
-
- Have a thin phosphor plate (0.5 to 2 mm thick)
- Have detectors parallel to the incident
- x-ray
- Each crystal needle is very fine between
- 5 and 20 µm in width
- The edges of the crystals act as light
- guides, the detector can either be on top of the crystals or below
-
- Selenium is a metalloid material,
- exhibiting both insulator and conductor properties depending on it’s state – it
- has several allotropes (cf
- sulphur or carbon).
- The
- selenium-based technology uses an amorphous selenium-coated
- thin-film-transistor (TFT) array to capture and convert X-ray energy directly
- into digital signals.
- Incident X-rays directly generate electron-hole
- pairs in the selenium layer
- via photoelectric absorption.
- These charges are collected by individual
- storage capacitors associated with each detector element for readout by
- customised electronics within the array.
- A Bias voltage is applied across the detector
- structure during the exposure
- When the exposure is terminated the
- individual detector elements are read, which generates the x-ray image

- Bildunterschrift: g) Low Potential, electrons can ‘easily’ accumulate on adjacent capacitor plates h) Thicker photoconductor means more electron/hole pairs generated, but still low potential so see above i) High field constrains electrons to discrete capacitor plates, thick photoconductor ensures plenty on electron/hole pairs
-
- Software
- converts to appropriate grey scale pixels according to voltage (proportional
- to intensity of incident x-ray photons)
- Higher
- spatial
- resolution - mammography applications down to 50 microns
- Direct
- conversion of x-rays to an electrical charge
- Can
- have in excess of 7 million detectors on a 40cm x 40cm plate.
- Images
- are available within 3-8 seconds
- DDR Advantages
- No processing time and almost immediate
- image acquisition
- Excellent image quality
- Pre readout info attributes appropriate
- image algorithm
- Immediate transfer to PACS / ER / Review
- terminal
- DDR Disadvantages
- More expensive than CR
- The imaging sensor is more expensive to
- replace than an imaging plate or cassette (of the order of ~ £20,000 each c/w
- ~£1-2000 each for 35 x 43 CR plate)
- Cannot usually
- retrofit to existing x-ray equipment – but some companies offer stand alone
- solutions
-
- Dose
- Slightly
- reduced with DR – depending on whose literature you read, upto 50%
- dose saving
- Resolution
- CR and DR are similar, DR is better for very
- high resolution applications
- Speed
- / throughput
- DDR
- is faster as there is no intermediate steps before next image acquisition
- Can
- use ‘tomosynthesis’ –
- repeated acquisition in plane at different projection angle to enhance lesion
- detection – lower dose than standard multi-planar projection
- Cost
- CR is
- cheaper than DDR

- Bildunterschrift: Amorphous selenium–based direct conversion DR systems. (a) Drawing illustrates a selenium drum–based system. A rotating selenium-dotted drum with a positive electrical surface charge is exposed to x-rays. Alteration of the charge pattern of the drum surface is proportional to the incident x-rays. The charge pattern is then converted into a digital image by an analog-to-digital (A/D) converter. (b) Drawing illustrates a selenium-based flat-panel detector system. Incident x-ray energy is directly converted into electrical charges within the fixed photo-conductor layer and read out by a linked TFT array beneath the detective layer.
-
- These are variants to flat panel
- detectors – used in cameras, video cameras, telescopes, and some mobile
- phones*.
- Electrons are liberated through the same
- process as CR, but using visible light instead of x-ray photons.
- The principle difference is that the
- detector is itself an integrated circuit made of crystalline silicon.
- The pixel electronics are etched onto the
- surface of the chip.
- The images produced in this way are of
- high quality.
- They are used in video fluoroscopy and
- cine radiography applications, and for very small field radiography such as
- dental applications.
- Quip about CMOS detectors and infra red
- detector (remote control)
- cheaper cameras use CMOS detector chips – lower resolution lower cost
- CCD
- cameras – higher resolution, higher cost