mass spec

Karten in diesem Stapel (38)
- 1what do mass spectrometers measureatomic mass unit per charge
- 2definition of atomic mass unita physical constant equal to one-twelfth of the mass of an unbound atom of carbon-12
- 3MS what does fragmentation help to identify - and what has an impact on the fragment patternfunctional groups in a molecule
- structure
the ionisation pattern has an impact on the fragment pattern - 4MASS SPEC info- used to determine elements
- highly versatile technique
- detects the mass of analyte (or fragment) with high accuracy - 5flow chart of the basic componence of a MASS SPECTROMETER1. sample inlet
2. vacuum system
3. ionisation method
4. mass analyser
5. detector
6. signal processor and data interpretation - 61. SAMPLE INLET what are the ways samples of different phases can be introduced without breaking the vacuumgas/volatiles - LEAK valve (thin capillary)
liquids - electrospray
solids - temperature programmed lance / SIMS - 73. IONIZATION METHOD hard ionization basic info- energetic
- causes lots of fragmentation - 83. IONIZATION METHOD soft ionization basic info- lower energy
- minimises fragmentation (simpler molecular spectrum)
- can identify parent mass - 9electron impact ionization basic info- gas phase
- lots of fragmentation - 10electron impact ionization what happensheated tungsten filament is used to produce electrons that bombard the analyte molecules, this causes ionization through loss of an electron
M + e- → M(*)+ • + 2e-
M+ typically has extra energy and undergoes further decomposition/fragmentation
M*+ → X+ + Y• - 11electron impact ionization what is the first step of the ionization method1. ionisation
thermionic emission - electrons released from a hot filament
electrons accelerated by 70 eV towards the trap (anode) due to potential difference between the filament and the trap
electrons collide with the analyte vapour, which undergoes ionization and fragmentation - 12electron impact ionization what is the second step of the ionization method2. generating an ion beam
positive ions = repelled by repeller
held at a positive potential
(4kV, energy 4k eV)
ions gain kinetic energy and move toward the mass analyser - 13what is the equation that underpins electron impact ionizationvelocity = √((2E/m) )
VELOCITY DEPENDS ON MASS - 144 MASS ANALYSER magnetic sector: speed basis costspeed = slow
basis = acceleration in magnetic field
cost = moderate - 154 MASS ANALYSER explain magnetic deflectionion beam constitutes a current
currents intact with magnetic fields, producing a perpendicular force - HENCE CIRCULAR PATH - 164 MASS ANALYSER what are the equations involved in magnetic deflectionforce applied by the field:
F = qvB
force to move a mass, m, in a circular motion with radius, r:
F = mv^2/r
circular motion for our ion in a magnetic field:
Br = mv/q - 174 MASS ANALYSER what happens at the magnetic sector1 accelerated ions passed into the magnetic sector
2 the mass spectrum is recorded by "scanning out" individual ions by varying the magnetic field, B
(or acceleration potential, V, hence ions momentum (m x v))
3 at a specific B, only certain ions aquire the correct circular velocity to reach the detector - 185 DETECTOR faraday cup info and steps- no internal amplifications
- fairly common
1. charged ions pass into the collection cup
2. upon impact they are neutralised and pass a charge to the cup
3. this induces a current, which is measures - 195 DETECTOR electron multiplier info and steps- most common detector in MS
- series of electrodes with large voltage difference
1. ions collide with an electrode which releases electrons
2. these electrons cascade down a series of electrodes - amplifying the signal up to 20 DYNODES (signal increased by a factor of 10^7) - 20fragmentation reactions: what is alpha cleavageunpaired electrons pair up depending on electron donating ability

- 21fragmentation reactions: sigma bond cleavagecarbocation is cleaved at the point of greatest substitution

- 22fragmentation reactions: inductive cleavageelectronegative hetero cation gains a pair of electrons to become neutral

- 23fragmentation pattern: EVEN ELECTRON RULEelectron ionization tends to form a radical cation
typically forms an even electron cation and a neutral radical - 24chemical ionization: phase fragmentationphase = gas
fragmentation = some - 25how does chemical ionization work1. ionization source chamber is filled with a reagent gas at a pressure ~1 mbar (HIGH = LOTS OF COLLISIONS)
2. energetic electrons (100-200 eV) ionise the reagent gas
3. the ionized reagent is used to ionize the analyte - 26chemical ionization what type of compounds is it good for what type of compound is excluded (and the outcome of this) what is the process very compatible with- good for volatile organic compounds
- excludes inorganics = very selective
very compatible with water-based atmospheric pressure samples - 27electrospray ionization - phase - fragmentationphase = liquid
fragmentation = very little - 28how does electrospray ionization work? 7 steps1. solution of analyte contains ions
2. liquid sample is passed through a fine nozzle
3. tip has a very high positive potential, drags solution out of the spray nozzle
4. charged droplets break away - due to thier +ve charge, they move towards the cathde
5. as the droplets cross the space they lose solvent and get smaller, increases their charge
6. AT THE RAYLEIGH LIMIT, coloumb fission occurs and form smaller charged particles - continues until only bare charged ions reach the mass spec
7. back flow of nitrogen gas helps the desolvation process - 29electrospray ionisation what is this process used forlarge non-volatuke biomolecules and liquid based systems
- as samples are takem directly from solution - 30electrospray ionisation nozzle sizes1-10 um nanospray
>10 um standard esi (pumping to help spray) - 31electrospray ionisation pH control of sample solutionhas a strong effect on the ionisation efficiency and the distribution of the analyte ion charge states
- sample solutions for electrospray MS usually are buffered or have added acids to enhance and control the formation of ions - 324. MASS ANALYSER quadrupole mass analyzer: speed basis costspeed = fast
basis = passage through ac electric field
cost = moderate - 332. MASS ANALYSER how does a quadrupole mass analyser work1. two pairs of rod electrodes through which a radioactive frequency current is passed, the two are 180˚ out of phase
2. beam of analyte travels between the rods
3. ions pass through quadrupole with oscillating trajectory
4. at a specific RF frequency and power, only specific m/z units will reach the end of the quadrupole and detector - 34what axis do ions pass through the quadrupolez axis
- 35what does stringing quadrupoles together do?results in higher resolution
- 364. MASS ANALYSER ion trap mass analyser speed basis costspeed = fast
basis = orbit in quadrupole
cost = moderate - 37how does an ion trap mass analyser workquadrupole where sample ions of certain m/z will have stable trajectory between the rods
positively charged plates at either end of analyser which repel positive ions
- ions passed back and forth = ion trap - 38how are ions confined in an ion trap mass analyserions held at a certain m/z
confining and anti-confininf directions are switched at a rate faster than it takes the particle to escape - confined in a figure of 8 pattern