  1. 1Why must a gamma camera be "peaked" every morning before use?
    1. Ato ensure a proper correction factor is applied to each pixel
    2. Bto ensure there is no damage to the camera system
    3. Cto ensure the camera is performing consistently day-to-day
    4. Dto ensure a proper correction factor is applied to each PMT and the spaces between PMTs
    5. Eto ensure the PHA knows which energies to accept in imaging studies
  2. 2Which is the correct order of events that occur when a gamma ray is emitted from the patient?

    1. A count is added at the x,y pixel location on the image matrix
    2. The photocathode produces electrons which accelerate and multiply through the dynodes
    3. The scintillation travels through the light pipe to the photocathode
    4. The pulse travels through the PHA
    5. A gamma ray passes through the collimator
    6. The position circuit determines the x & y signal locations
    7. A gamma ray interacts with the NaI(Tl) crystal to produce scintillation
    8. The amplifier manipulates the Z-pulse
    1. A5 > 7 > 3 > 2 > 6 > 8 > 4 > 1
    2. B7 > 5 > 3 > 2 > 8 > 6 > 4 > 1
    3. C5 > 7 > 4 > 6 > 1 > 3 > 2 > 8
    4. D5 > 7 > 3 > 2 > 6 > 1 > 8 > 4
    5. E5 > 7 > 3 > 2 > 6 > 4 > 8 > 1
  3. 3A nuclear medicine department receives a new gamma camera. The new camera has a larger number of smaller photomultiplier tubes than the older cameras currently in use. Which imaging characteristic will be most improved with the new camera?
    1. Auniformity
    2. Benergy resolution
    3. Cspatial resolution
    4. Dsensitivity
    5. Eminimum detectable activity
  4. 4 When determining the X and Y coordinates of a gamma event, the positional signals are divided by the energy signal (Z) to normalize the signal. Failure to normalize the positional signal may have what result on the resulting image?
    1. ATrue events may not be accepted by the pulse-height analyzer
    2. BAn event may be mapped to the wrong location on the image matrix
    3. CEdge packing will be increased
    4. DA ring artifact may occur
  5. 5Which of the following is the correct definition of spatial resolution as it relates to conventional nuclear medicine imaging?
    1. Athe ability of a gamma camera to produce a uniform image in response to a uniform source of gamma rays
    2. Bthe ability of a gamma camera to use the gamma rays available to it
    3. Cthe ability of a gamma camera to reproduce the details of a nonuniform source of gamma rays
    4. Dthe ability of a gamma camera to accurately reflect the different energies of a radionuclide source
    5. Ethe ability of a gamma camera to correct for attenuation in an object
  6. 6A static image is acquired with a 128x128 matrix. The minimum pixel value is 120 and the maximum pixel value is 18540. If using an 8-bit LUT array, what shade of grey will a pixel with a value of 18000 be assigned?
    1. A248
    2. B250
    3. C249
    4. D247
    5. E245
  7. 7To obtain high resolution images of a small organ, gland, or joint, which collimator would be most useful?
    1. Adiverging
    2. Bflat field
    3. Chigh-energy diverging
    4. Dlow-energy parallel-hole
    5. Epinhole
  8. 8A technologist changes the collimator on a gamma camera from a LEAP to a LEHR collimator. If the same number of counts is acquired, how will the acquisition time change when the high-resolution collimator is used?
    1. Athe acquisition time will be the same with both collimators
    2. Bthe acquisition time will decrease
    3. Cthe acquisition time will increase
    4. Dthe effect on acquisition time is unpredictable
    5. Ethe acquisition time is determined by the intrinsic sensitivity of the instrument
  9. 9For a parallel hole collimator, the best resolution is at:
    1. Athe collimator surface
    2. B10 cm
    3. C15 cm
    4. D20 cm
    5. Ethe focal distance
  10. 10Which of the following will increase the resolution of a "spot view" bone image obtained with a parallel hole collimator?
    1. Amoving the camera closer to the patient's body
    2. Bincreasing the PHA window width
    3. Cusing a high sensitivity collimator instead of a LEAP collimator
    4. Dusing fewer shades of grey to display the image
    5. Eusing a slant hole collimator