Computed Radiography & Digital Radiography

Rutendo Chingomb Diagnostic Radiography 1 Inglés 17 diapositivas

Diapositivas del conjunto (17)

  1. CR and DR

    • Image
    • Receptor:
    • –Film
    • screen
    • –Computed
    • Radiography
    • –Direct
    • Digital Radiography (DDR/DR)
    • The
    • computer processing means that systems have a far greater latitude compared to
    • film/screen. But
    • knowing what happens when you set/manipulate exposure factors still matters
    • Increasing kVp also
    • increases beam quantity to a
    • certain extent: because the electrons with more energy are more efficient at
    • removing K-shell electrons. Intensity
    • ∝ kVp2
    • CR has a tube potential (kVp) that determines the subject contrast in the resultant image - this is connected to K- edge of the receptors
    • Anatomical tissues vary in absorption and
    • transmission create a range of dark and light areas
    • Remember it is the degree of
    • photoelectric absorption that determines the range of greys visible on the
    • radiographic image. This degree of absorption is strongly dependent on both the
    • density and inversely proportional to beam energy
    • This is what creates the radiographic
    • image. Since it is the interaction of the outer electrons that usually
    • influence the degree of absorption, transmission, or scatter, the radiographic
    • can be thought of as a map of electron density
    • And whilst CR and DR has a wide exposure latitude if there isn't enough information to make a diagnostic image then there isn't enough!
  2. Exposure Latitude

    • Formally
    • it is the tonal range generated within an exposure.
    • Sometimes
    • called a characteristic curve, that has axes of transmitted light (ie
    • range of greys very light to very dark), and exposure (logarithmic
    • Sometimes
    • called the dynamic range
    • In
    • digital imaging, changing kVp affects the amount of
    • radiation reaching the image receptor and
    • affects subject
    • contrast (differential attenuation or inherent contrast)
    • In
    • digital imaging, brightness and image contrast are primary
    • controlled during computer processing – but you still need to be aware of the
    • effect of kVp on
    • subject (inherent) contrast
    • Exposure Latitude
    • Pie de imagen: Speed of ‘film’ is related to the gradient – a faster film has a steeper gradient, historically because the crystals which made up the film were bigger, requiring less exposure to oxidise to silver. A slower film therefore had smaller crystals requiring more exposure per unit time to generate the same degree of oxidation
  3. Diapositiva 3

    • Film screen vs CR
    • CR
    • and DR are digital imaging modalities. Both CR and DR are relative newcomers to
    • x-ray imaging
    • Film-screen
    • uses a cassette similar to the cassette used in CR
    • Film-screen
    • uses a processor with chemicals where as CR uses a digital processor
    • In
    • CR the film and intensifying screen is replaced with a phosphor plate
    • CR equipment:
    • Cassette
    • and phosphor plate, Laser
    • Image Reader, PACS
    • CR Advantages
    • –A
    • digital image is generated
    • –Ability
    • to retrofit to existing radiography equipment
    • –Mobile
    • radiography is easily accomplished
    • –Excellent
    • image quality
    • –Initially
    • less expensive than DR
    • CR disadvantages
    • –Still
    • have to use an imaging plate
    • –No
    • real time saving benefit over traditional radiography
    • –Need
    • to purchase an imaging reader
    • Imaging Plate
    • Average
    • life cycle over 10000 uses
    • Two
    • year warranty
    • Average
    • cost £ 1000 each
    • Flexible
    • Durable
    • High
    • Absorption efficiency - high
    • % of incoming x-rays effectively contribute to information capture, therefore
    • enabling x-ray dose reduction
    • Smooth
    • Shiny surface
    • Correct
    • & Incorrect sides? TUBE side
    • 35 x 43 cm, 35 x 35 cm, 24 x 30 cm, 18 x 24 cm, 15 x 30 cm
    • Storage: Upright ( NOT flat) to prevent pressure on the Phosphor plate. Must not be left unprotected in the x-ray room, sensitive to scatter.
    • Cleaning: Every TWO weeks and erased. If not there is an increased risk of becoming stuck in the digitizer.
    • Cassette Erasure –Erased ONCE a week at least, UHL it is done first thing every day. Reduce risk of ghost image.
  4. Phosphor Plate

    • Phosphor plate: Photo-stimulable storage phosphor plate, that stores the
    • latent image, is a re-usable detector.
    • Two
    • part process:
    • 1.converting incoming x-rays to light
    • photons - image capture
    • 2.converting light into electrical signal - image readout
    • Laser Image Reader: Scanning mechanism to
    • extract the latent image from the plate.
    • Transmitted to network for
    • review/reporting etc
    • Through the PACS system
    • Phosphor
    • plate is analogous to film – it is sensitive to x-ray irradiation, whereas film
    • is not; BUT to demonstrate what happens during an x-ray exposure instead of an intensifying screen – a scintilator
    • which converts x-ray to visible light which exposes a film, the light is
    • trapped within the plate to be read out later
    • Phosphor Plate
  5. Diapositiva 5

    • Computed Radiography & Digital Radiography
    • Protective layer - thin
    • & transparent, tough
    • and clear. Protects the phosphor layer during handling
    • from damage.
    • Phosphor layer - active
    • component, barium fluorohalide
    • crystals doped with europium, contains
    • colour centres or F centres, which trap electrons during an exposure.
    • Support
    • layer - semi
    • rigid, give
    • strength to the plate
    • Conductor
    • layer - absorbs
    • and removes static electricity
    • Light
    • shield layer - reflective
    • layer, sends
    • light in forward direction when released in the reader, ensuring as much light
    • as possible is recorded
    • Backing
    • layer - protects the back of the cassette.
  6. Latent Image Acquisition

    • •Exit
    • photons are absorbed by the atoms of the phosphor and europium
    • •This
    • is by the photoelectric effect
    • Phosphor plate - image capture
    • The absorbed energy excites the electrons
    • They are raised to a higher energy state
    • Here they become stored/trapped in colour (F) centres
    • This is what forms the latent image
    • The number of electrons trapped in the F centres is directly related to X-ray beam intensity
    • This energy ( latent image ) can be stored for several hours
    • However the image will lose about 25%of its energy in 7-8 hours
    • The latent image is processed by loading the cassette into an image reader.
    • Image reader unit (digitiser)
    • The cassette is placed into the reader
    • unit
    • The imaging plate is extracted and
    • scanned
    • The image plate is scanned with a finely
    • focussed helium-neon laser
    • beam.
    • The laser releases the trapped electrons
    • and they return to a lower energy state.
    • This releases energy as visible light.
  7. Digitiser

    • Digitiser
    • The
    • varying electrical signals are analogue, converted by an ADC to a digital image.
    • Once
    • the plate is read it is erased to remove all trace of the latent image
    • By
    • exposing the plate with intense white light to release any trapped electrons.
    • Imaging
    • plates are sensitive to `Scatter radiation` and should be erased once a week to
    • prevent background signal build up. The readers have an erase mode
    • The scanning of the plate results in a
    • continuous pattern of light intensities sent to a photomultiplier tube (which
    • is a photo detector)
    • Photomultiplier tube
    • collects, amplifies and coverts the light to an electrical signal
    • The electrical signal is proportional to
    • the range of energies stored in the image plate.
  8. Matrix

    • A
    • digital image is made up of a 2D array of numbers called a Matrix
    • Bit depth - A
    • descriptor of the grey scale value attributed to each picture element
    • Pixels are the smallest component of a matrix
    • The density of each Pixel ( picture
    • element) has to be a whole number. Each pixel has a bit depth or a number of
    • bits which controls the brightness of the pixel – 4bit, 8 bit, 10 bit or 12 bit
    • The greater the bit depth the more the
    • precise digitalization of the analogue signal, and the greater the various
    • shades of grey
    • This is turn improves its contrast
    • resolution - a previous slide suggested
    • that 1024 discrete shades of grey can be displayed - but we should consider actually how shades of
    • grey can the eye distinguish under optimum image display conditions?
    • It
    • is now established that the human eye can discriminate 'between 700 and 900
    • simultaneous shades of gray for
    • the available luminance range of current medical displays and in optimal
    • conditions.' It is presently felt that it is of 'no use to simultaneously
    • display more than 10 bits of gray (1,024 gray shades) because this already exceeds the
    • capabilities of the human visual system
    • Larger the matrix size, greater the
    • number of pixels, better the image quality
  9. What next

    • Now we have an image, what happens next?
    • PACS – Picture Archiving and
    • Communication System
    • Images are captured by a “ Image Server
    • ”. PACS does not generate any images, simply handles the data.
    • The Server is the primary point of entry. It
    • receives digital signals from all radiology modalities, CT, MR, NM, C-arm, DSA,
    • digitisers…
    • The language of medical digital imaging is
    • called “DICOM”.
    • DICOM allows various components to communicate
    • with each other.
    • In PACS these images are captured as data
    • frames and routed through a server to digital display monitor.
    • For more information please see resources
    • on blackboard
  10. CR vs DR

    • CR imaging screens are ‘doped’ usually
    • Barium Fluorohalide;
    • with a small quantity of Europium.
    • The addition of Eu
    • causes defects within the crystal lattice which allows electrons to become
    • trapped more efficiently (F-Centre)
    • Incident x-ray is absorbed by the BaFH:Eu
    • phosphor – which oxidises the Eu atoms, the electrons liberated by this
    • process become trapped in a higher energy metastable
    • state
    • This is the ‘latent’
    • image
    • The number of trapped electrons is
    • proportional to the intensity of the incident x-ray beam
    • Flat panel: Caesium Iodide or Selenium
    • detector ( FPD)
    • Absorbs x-rays directly and converts into
    • a digital signal in 4 - 8 seconds
    • Indirect flat panel detectors use an intensifying screen or scintillator to convert x-rays to visible light, and then to a photodetector to produce a visible image
    • Direct flat panel detectors use incident
    • x-rays to generate free electrons which generate the ‘latent’ image directly.
    • Selenium is commonly used as a
    • photoconductor
    • A large voltage difference is applied
    • across the array at the same time, which causes the electrons to move to their
    • detectors.
  11. Indirect Digital Radiography

    • Indirect Digital Radiography
    • Caesium Iodide doped detectors are used as indirect flat panel detectors – there is a intermediate step in the generation of a digital signal.
    • Incident x-ray photons interact via photoelectric absorption
    • This leads to scintillation
    • The light is detected by a photodiode
    • The intensity of the light is directly
    • proportional to the number of incident x-ray photons
  12. Blurring and spatial resolution

    • Blurring and spatial resolution
    • a)Internal
    • reflection and diffusion processes
    • b) the electron can wander about ionising
    • as it goes and losing energy, or
    • c)Can be absorbed by another atom leading
    • to another ionisation type event, with corresponding re-emission of light. This
    • would simulate x-ray incidence additional to the primary point
    • d)How can you tell which is which – You
    • cant, since the number of interactions is a phenomenal amount (259 * 6.023
    • x10exp23) ions present in 1 mole of CsI
    • e)Approx ten times the number of stars in the known
    • universe (1 x 10 exp 24)
    • Blurring and spatial resolution
    • Pie de imagen: d) Wide range of intensity e) Narrower range, but decreased amplitude –so less light available f) So structured phosphor is the way to go
  13. Solutions

    • Have a thin phosphor plate (0.5 to 2 mm thick)
    • Have detectors parallel to the incident
    • x-ray
    • Each crystal needle is very fine between
    • 5 and 20 µm in width
    • The edges of the crystals act as light
    • guides, the detector can either be on top of the crystals or below
  14. Direct Digital Radiography

    • Selenium is a metalloid material,
    • exhibiting both insulator and conductor properties depending on it’s state – it
    • has several allotropes (cf
    • sulphur or carbon).
    • The
    • selenium-based technology uses an amorphous selenium-coated
    • thin-film-transistor (TFT) array to capture and convert X-ray energy directly
    • into digital signals.
    • Incident X-rays directly generate electron-hole
    • pairs in the selenium layer
    • via photoelectric absorption.
    • These charges are collected by individual
    • storage capacitors associated with each detector element for readout by
    • customised electronics within the array.
    • A Bias voltage is applied across the detector
    • structure during the exposure
    • When the exposure is terminated the
    • individual detector elements are read, which generates the x-ray image
    • Direct Digital Radiography
    • Pie de imagen: g) Low Potential, electrons can ‘easily’ accumulate on adjacent capacitor plates h) Thicker photoconductor means more electron/hole pairs generated, but still low potential so see above i) High field constrains electrons to discrete capacitor plates, thick photoconductor ensures plenty on electron/hole pairs
  15. DDR image receptor

    • Software
    • converts to appropriate grey scale pixels according to voltage (proportional
    • to intensity of incident x-ray photons)
    • Higher
    • spatial
    • resolution - mammography applications down to 50 microns
    • Direct
    • conversion of x-rays to an electrical charge
    • Can
    • have in excess of 7 million detectors on a 40cm x 40cm plate.
    • Images
    • are available within 3-8 seconds
    • DDR Advantages
    • No processing time and almost immediate
    • image acquisition
    • Excellent image quality
    • Pre readout info attributes appropriate
    • image algorithm
    • Immediate transfer to PACS / ER / Review
    • terminal
    • DDR Disadvantages
    • More expensive than CR
    • The imaging sensor is more expensive to
    • replace than an imaging plate or cassette (of the order of ~ £20,000 each c/w
    • ~£1-2000 each for 35 x 43 CR plate)
    • Cannot usually
    • retrofit to existing x-ray equipment – but some companies offer stand alone
    • solutions
  16. CR vs DDR

    • Dose
    • Slightly
    • reduced with DR – depending on whose literature you read, upto 50%
    • dose saving
    • Resolution
    • CR and DR are similar, DR is better for very
    • high resolution applications
    • Speed
    • / throughput
    • DDR
    • is faster as there is no intermediate steps before next image acquisition
    • Can
    • use ‘tomosynthesis’ –
    • repeated acquisition in plane at different projection angle to enhance lesion
    • detection – lower dose than standard multi-planar projection
    • Cost
    • CR is
    • cheaper than DDR
    • CR vs DDR
    • Pie de imagen: Amorphous selenium–based direct conversion DR systems. (a) Drawing illustrates a selenium drum–based system. A rotating selenium-dotted drum with a positive electrical surface charge is exposed to x-rays. Alteration of the charge pattern of the drum surface is proportional to the incident x-rays. The charge pattern is then converted into a digital image by an analog-to-digital (A/D) converter. (b) Drawing illustrates a selenium-based flat-panel detector system. Incident x-ray energy is directly converted into electrical charges within the fixed photo-conductor layer and read out by a linked TFT array beneath the detective layer.
  17. Charge Coupled Devices (CCD)

    • These are variants to flat panel
    • detectors – used in cameras, video cameras, telescopes, and some mobile
    • phones*.
    • Electrons are liberated through the same
    • process as CR, but using visible light instead of x-ray photons.
    • The principle difference is that the
    • detector is itself an integrated circuit made of crystalline silicon.
    • The pixel electronics are etched onto the
    • surface of the chip.
    • The images produced in this way are of
    • high quality.
    • They are used in video fluoroscopy and
    • cine radiography applications, and for very small field radiography such as
    • dental applications.
    • Quip about CMOS detectors and infra red
    • detector (remote control)
    • cheaper cameras use CMOS detector chips – lower resolution lower cost
    • CCD
    • cameras – higher resolution, higher cost
    • Charge Coupled Devices (CCD)