Removal of gross soil.
Prevention of damage.
Prevention of cross contamination.
Keeping others safe.
Slide 2
Point-of-Use Preparation
Instrument preparation begins at the point of use.
Slide 3
Point-of-Use Preparation Guidelines
Remove gross soil.
Follow the manufacturer's instructions for point-of-use cleaning and preparation.
Separate reusable sharps from other instruments.
Separate reusable linen.
Remove disposable components.
Open hinged instruments, disassemble multi-part instruments and place them in trays. Heavy: bottom, Light: top, Cords, Scopes, Cameras: Very top.
Keep items together. (Sets, multi-part items)
Slide 4
Continued..
Keep instruments moist and prevent soil from drying on their surface. Use foams, gels, sprays, or a moist towel. Limit the time kept in a soak basin or solution to prevent instrument damage.
Empty fluids from contaminated device containers using facility protocols.
Remove, contain, or separate third-party reprocessing items and send to CS, per facility protocols.
For reusable instruments used in a case with CJD notify CS and Infection Prevention Control. Follow hospital procedures.
Notify CS: Items needing repair. Tag them to remove them from the system.
Notify CS: Items requiring turnover/turnaround.
Slide 5
Transport of Soiled Items
Transport soiled instruments in enclosed carts. (Preferred)
Open Carts: Must have a solid bottom/ cover cart during transport.
Don't place metal trays/containers on top of plastic trays/containers.
Don't use containers/carts transporting contaminated items to transport clean and sterile items. (Unless the cart is decontaminated between uses)
Slide 6
Education and Training
AAMI recomends practices for transporting contaminated items.
Procedures should also reflect OSHA regulations.
AORN provides additional information.