Gelöschter Nutzer 7 Fragen Aktualisiert am 25 Aug 2016

Quiz starten

Dieses Quiz enthält Fragetypen, die sich im gewohnten Player öffnen.

Fragen in diesem Quiz (7)

  1. 1Gas Chromatography (Absorption Type) NEED TO ADD THIS [blank_start]QUESTION BACK[blank_end]
    1. AQUESTION BACK
  2. 2Advantages: Analyzes [blank_start]many gases[blank_end] at once. Very accurate
    Disadvantages: Requires more [blank_start]time[blank_end] than other analyzers.
    Affected by: [blank_start]H2O[blank_end] and [blank_start]CO2[blank_end]
    Measures: [blank_start]Ne[blank_end], [blank_start]O2[blank_end], [blank_start]N2[blank_end], [blank_start]CO[blank_end]
    1. ACO
    2. BN2
    3. CO2
    4. DNe
    5. ECO2
    6. FH2O
    7. Gtime
    8. Hmany gases
  3. 3CO2/CO/CH4

    AKA: [blank_start]Infrared[blank_end] Absorption Analyzer
    Quick response time is suitable for [blank_start]DLCO[blank_end]
    Calibrate: to room air ([blank_start]0[blank_end]%) and known concentration (.[blank_start]3[blank_end]%) CO then re-zero and use [blank_start]gain[blank_end] to match read out to [blank_start]known[blank_end] concentration.
    Changes in altitude do [blank_start]NOT[blank_end] affect readings.
    1. AInfrared
    2. BDLCO
    3. C0
    4. D3
    5. Egain
    6. Fknown
    7. GNOT
  4. 4Helium analyzer

    Also Known as: [blank_start]Thermal[blank_end] Conductivity or [blank_start]Wheatstone[blank_end] Bridge
    Measures: Gas [blank_start]percentage[blank_end]
    Calibration: To room air ([blank_start]0[blank_end]%) and known ([blank_start]10[blank_end]%) then use gain to adjust to [blank_start]known[blank_end] amount.
    Do not use with [blank_start]flammable[blank_end] gases.
    Used with Helium dilution tests.
    1. AThermal
    2. BWheatstone
    3. Cpercentage
    4. D0
    5. E10
    6. Fknown
    7. Gflammable
  5. 5Mass Spectometry
    Function: Sample is drawn in by vacuum to the [blank_start]ionization[blank_end] chamber. Individual gases are [blank_start]separated[blank_end] and analyzed.
    Measures: Gas percentages
    Advantages: [blank_start]Rapid[blank_end] Response time, breath by breath analysis, multiple gas analysis
    Can analyze: [blank_start]He[blank_end], [blank_start]N2[blank_end], [blank_start]O2[blank_end], [blank_start]CO2[blank_end]
    Disadvantages: Expensive, large, high maintenance
    Most Suitable for: analysis of [blank_start]several[blank_end] patient at once rather than individual testing.
    1. Aionization
    2. Bseparated
    3. CRapid
    4. DHe
    5. EN2
    6. FO2
    7. GCO2
    8. Hseveral
  6. 6Nitrogen analyzer

    Also known as: [blank_start]Geissler[blank_end] Tube Ionizer
    Sample is pumped into [blank_start]ionization[blank_end] chamber where light is emitted and monitored by a phototube. Intensity of light is directly [blank_start]proportional[blank_end] to N2%
    Measures gas percentage
    Capable of breath by breath analysis
    Calibration: To zero with [blank_start]100[blank_end]% O2 and a [blank_start]known[blank_end] concentration.
    1. AGeissler
    2. Bionization
    3. Cproportional
    4. D100
    5. Eknown
  7. 7O2 Analyzers
    Galvanic [blank_start]Fuel[blank_end] Cell Analyzer: Measures partial [blank_start]pressure[blank_end] and displays [blank_start]FiO2[blank_end] as %.
    If unable to calibrate you should replace the [blank_start]fuel[blank_end] cell.
    Calibration during exercise testing: Low [blank_start]12[blank_end]-[blank_start]14[blank_end]% and high [blank_start]21[blank_end]%

    Poloragraphic Analyzer
    Similar to galvanic but uses a [blank_start]battery[blank_end]
    Analogous to the [blank_start]Clark[blank_end] electrode
    Calibration is to [blank_start]21[blank_end]% and [blank_start]100[blank_end]%
    1. AFuel
    2. Bpressure
    3. CFiO2
    4. Dfuel
    5. E12
    6. F14
    7. G21
    8. H21
    9. I100
    10. JClark
    11. Kbattery

Informationen

Beschreibung
Der Nutzer hat seine Fachbereichs-Information gelöscht. Quiz am Gas Anaylzers, erstellt von Gelöschter Nutzer am 19/08/2016.
Autor
Gelöschter Nutzer
Sprache
English
Aufrufe
40
Verlauf
Erstellt am 19 Aug 2016 · aktualisiert am 25 Aug 2016

Feedback

Worum geht es?

Wie gefällt dir das neue Design?

Dein Eindruck