ABG Analyzers

Beschreibung

Equipment (Gas Analyzers) Quiz on ABG Analyzers, created by LeeAnna Shepherd on 08/15/2016.
LeeAnna Shepherd
Quiz von LeeAnna Shepherd, aktualisiert more than 1 year ago
LeeAnna Shepherd
Erstellt von LeeAnna Shepherd vor mehr als 7 Jahre
15
0

Zusammenfassung der Ressource

Frage 1

Frage
ABG Analyzers Directly measures [blank_start]PCO2[blank_end] - Severinghaus, [blank_start]PO2[blank_end] - Clark, pH - [blank_start]Sanz[blank_end] HCO3, SaO2 and other values are calculated Formula for read out: (Barometric Pressure - [blank_start]47[blank_end]) x Fractional Concentration of [blank_start]gas[blank_end]
Antworten
  • PCO2
  • PO2
  • Sanz
  • 47
  • gas

Frage 2

Frage
Severinhaus Electrode Measures [blank_start]PCO2[blank_end] [blank_start]Glass[blank_end] electrode in [blank_start]lucite[blank_end] jacket Includes nylon spacer and [blank_start]teflon[blank_end] membrane
Antworten
  • PCO2
  • Glass
  • lucite
  • teflon

Frage 3

Frage
PCO2 Calibration: Cylinders containing known concentrations of gas: Low: [blank_start]5[blank_end]% balance High: [blank_start]10[blank_end]% slope
Antworten
  • 5
  • 10

Frage 4

Frage
Clark Electrode Measures [blank_start]PO2[blank_end] Platinum cathode, silver-silver chloride anode, polypropolene membrane (gas [blank_start]permeable[blank_end]) Using [blank_start]mylar[blank_end] membrane results in insufficient O2 diffusion Cleaning top with [blank_start]pumice[blank_end] periodically to clean [blank_start]protein[blank_end]
Antworten
  • PO2
  • permeable
  • mylar
  • pumice
  • protein

Frage 5

Frage
PO2 Calibration: Cylinders containing known concentrations of gas: Low: [blank_start]0[blank_end]% balance High: [blank_start]12[blank_end]% or [blank_start]20[blank_end]% slope
Antworten
  • 0
  • 12
  • 20

Frage 6

Frage
Sanz Electrode Measures PH Reference Electrode Contains: [blank_start]KCL[blank_end] solution with known PH, has contsant [blank_start]electrical[blank_end] potential. Measuring electrode contains: [blank_start]Buffer[blank_end] Solution, silver-silver chloride wire and pH sensitive glass capillary tube Introduction of sample causes electrical [blank_start]potential[blank_end] to develop. Voltage difference is proportional to PH [blank_start]difference[blank_end]
Antworten
  • KCL
  • electrical
  • Buffer
  • potential
  • difference

Frage 7

Frage
Sanz Electrode Calibration: Low pH solution: [blank_start]6.84[blank_end] - If displayed value is different than the known sample, adjust with the [blank_start]balance[blank_end] to known value. High pH solution: [blank_start]7.384[blank_end] - If displayed value is different than the known sample, adjust with the [blank_start]slope[blank_end] control to known value.
Antworten
  • 6.84
  • 7.384
  • balance
  • slope

Frage 8

Frage
Tonomotry Used for quality [blank_start]control[blank_end] with ABG analyzers Allows precise gas mixtures to be [blank_start]equilibrated[blank_end] with buffer solution or whole blood samples Best method for precise control of [blank_start]Clark[blank_end] electrode Vert complex - not widely used
Antworten
  • control
  • Clark
  • equilibrated
Zusammenfassung anzeigen Zusammenfassung ausblenden

ähnlicher Inhalt

Laboratory Tools and Equipment
Mark Anthony Sy
Identify common tools and equipment
T Andrews
Materials / Equipment for Badminton
Jonna Jean Barbosa
Group B, equipment project
Zoe Bishop
Equipment List
francesca.clover
TOOLS AND EQUIPMENT
Anett M
Tools and equipment
Karl Laidna
TRAVEL EQUIPMENT
Julita Hulewicz
Equipment Used and Purpose
Becca Westwell
Gas Anaylzers
LeeAnna Shepherd