CH 15: Low-Temperature Sterilization

Javier Serafin Central Service Technical Manual 8th Edition CRCST Inglés 37 diapositivas

Diapositivas del conjunto (37)

  1. Introduction

    • Ethylene Oxide (EtO)
    • Hydroden Peroxide (H2O2)
    • Ozone (O3)
  2. Low Temperature Sterilization Requirements

    • Efficacy (Effectiveness): Provides the minimum-required Sterility Assurance Level (SAL).
    • Safety: No toxic residues on packaging/devices upon completion of sterilization cycles.
    • Exposure Monitoring.
    • Sterilization Performance Monitoring.
    • Penetration.
    • Material Compatibility.
    • Adaptability.
    • Approval.
  3. Efficacy

    • FDA regulated.
    • FDA/CDC outlines the SAL for each sterilization method.
  4. Toxicity Standards: Low-Temp Sterilants

    • PEL: Permissible Exposure Limits; parts per million (ppm). By OSHA.
    • TWA: Time Weighted Average (over an 8 hour day). By OSHA.
    • IDLH: Immediately Dangerous to Life or Health. National Institute for Occupational Safety and Health (NIOSH) established.
    • EtO: OSHA PEL/TWA (1.0 ppm), NIOSH IDLH limit (800 ppm). Monitoring required.
    • H2O2: OSHA PEL/TWA (1.0 ppm), NIOSH IDLH limit (75 ppm). Monitoring is not required.
    • O3: OSHA PEL/TWA (0.1 ppm), NIOSH IDLH limit (5 ppm). Monitoring is not required.
  5. Penetration

    • Eto inactivates microbes (alkylation).
    • H2O2/O3 destroys microbes (oxidation).
  6. Ethylene Oxide

    • For heat and moisture-sensitive medical devices.
    • High vapor pressure and a low boiling point of 51.3 F
    • EtO cartridges are placed inside chambers of less than 10 cubic feet. The cartridge is punctured under vacuum.
    • Sterilization Cycle 5 Stages: Preconditioning and humidification, gas introduction, exposure, evacuation and air washes.
    • Cycle: About 2 1/2 hours, excluding aeration time.
    • After sterilization items must go through an aeration process (removes residual EtO before the chamber can be unlocked for item removal).
  7. Safety

    • OSHA classifies EtO as a carcinogen and reproductive hazard.
    • EtO Sterilizers: Located in a well ventilated area/ air exchange rate of at least 10 changes per hour/ in negative-pressure rooms.
  8. EtO Sterilization Process Parameters

    • Gas Concentration: 45 to 1,200 mg/l
    • Temperature: 98 F to 145 F
    • Relative Humidity: 40% to 80%
    • Exposure Time: 1 to 6 hours
    • Aeration Time: 8 to 12 hours at 122 F to 140 F
  9. Exposure Monitoring

    • Devices affixed to the employee's clothing/breathing zone (within 1 foot of the person's nose) tests airborne EtO concentrations.
  10. Materials Compatibility

    • EtO should not be used to sterilize: 1) Liquids, 2) Devices with energy sources (sparks are created), 3) Leather Items (chemicals used in the tanning process will combine with EtO and form chlorohydrin)
  11. Packaging

    • EtO paper/plastic or Tyvek peel pouches, approved fabric wrappers, medical crepe paper, polypropylene and most container systems.
    • Never use aluminium foil, cellophane, nylon films, polyester, PVC (plastic) films and styrofoam.
  12. Aeration

    • Vacuum pulse remove EtO.
    • Warm air circulates to remove residuals from instruments. Minimum 8 hours at 140 F and 12 hours at 122 F.
  13. Sterilizer Performance Monitors

    • Bacillus atrophaeus: Microorganism of choice for EtO.
  14. Hydrogen Peroxide Systems

    • Hydrogen Peroxide Gas Plasma: Oxidative Process. Safer than EtO, rapid cycle times. Byproducts of the cycle (water and oxygen) are nontoxic = no aeration phase.
    • Gas plasma are highly ionized gases.
  15. Efficacy

    • H2O2 Gas Plasma: Hydrogen Peroxide solution from 59% to 95% is proven effective at killing microorganisms.
  16. Penetration

    • H2O2 gas plasma sterilizers sterilize some single channel flexible endoscopes, cameras, rigid endoscopes, light cords, batteries and power drills.
  17. Types of H2O2 Gas Plasma Systems

    • Compact Systems: 28-38 min processing times.
    • Large Capacity Systems: 75 min cycle times.
  18. Sterilization Cycle/Process Parameters

    • Phases of hydrogen peroxide gas plasma.
    • 1)Vacuum: The load is heated. Removes remaining water as it evaporates.
    • 2)Injection: H2O2 into chamber.
    • 3)Diffusion: Drives H2O2 vapor into small crevices/lumens of devices.
    • 4)Plasma: A vacuum decreases the pressure/radio frequency (RF) energy is radiated from an electrode screen. RF energy ionizes the H2O2 creating H2O2 gas plasma.
    • 5)Vent: No aeration. Use instruments after the 24-75 min sterilization cycle.
  19. H2O2 Gas Plasma:Process Parameters

    • Time: 24-75 mins.
    • Cycle Temperature: Less than 131 F
    • Hydrogen Peroxide Concentration: 59% to 95%
    • Vacuum Level: Automatically controlled.
    • Plasma: Automatically controlled.
  20. Safety

    • H2O2 liquid irritates skin/damages eyes.
    • In vapor phase: Irritates eyes,nose, throat, and lungs.
    • H2O2 is packaged in sealed cassettes. Chemical leak indicators on them change from yellow or white to red.
    • After 5 sterilization cycles the cassette is automatically ejected into a collection box for safe disposal.
    • Latex, vinyl (PVC), or nitrile gloves are worn to remove items from a cancelled cycle.
  21. Materials Compatibility

    • H2O2 gas plasma is not compatible with:
    • 1) Liquids and powders.
    • 2) Any material that absorbs liquid.
    • 3) Items that contain cellulose, such as cotton, paper or cardboard, linens, huck towels, gauze sponges or any item containing wood pulp.
  22. Packaging

    • Tvek pouches are compatible.
    • Don't use cellulose-containing packaging such as paper/plastic pouches.
  23. Sterilizer Performance Monitors

    • Geobacillus Stearothermophilus: Microorganism of choice for H2O2 gas plasma.
    • AORN Guidelines: Daily biological monitoring.
  24. Vaporized Hydrogen Peroxide

    • "VHP": used since 2007.
    • 59% hydrogen peroxide solution.
    • Injected 4 times during each sterilization cycle. After the 4th injection period, the load is automatically aerated in the sterilizer. VHP is converted to water and oxygen.
  25. Types of Vaporized H2O2 Systems

    • Single Preprogrammed 55 Min Cycle: Lumened and nonlumened instruments and devices.
    • Two Preprogrammed Cycles: 28 min (non lumened instruments) and 55 min (instruments with lumens/non stainless steel mated surfaces)
  26. Sterilization Cycle Process Parameters

    • Phases of VHP systems include:
    • 1) Conditioning: Removes air/excess moisture then recharges with dry sterile air.
    • 2) Leak Test: Vacuum is held to ensure a leak tight chamber.
    • 3) Sterilization: 4 injection pulses, each followed by a hold period.
    • 4) Aeration: Automatic.
  27. Vaporized H2O2: Process Parameters

    • Cycle Tempeerature: Less than 122 F
    • H2O2 Concentration: 59%
    • Time: 28 to 55 mins.
  28. Safety

    • Concentrated H2O2 is corrosive to skin, eyese, nose, throat, lungs, and gastrointestinal tract.
    • Wear latex, vinyl (PVC), or nitrile gloves when removing items from a cancelled load.
  29. Materials Compatibility

    • Vaporized H2O2 sterilization is compatible with telescopes, cameras, light cables. batteries and many lumened devices.
    • The VHP system is not intended to process liquids, linens, powders, or any cellulose materials.
  30. Packaging

    • Polywrap, nonwoven sterilization packaging made of 100% polypropylene, Tyvek.
  31. Sterilizer

    • Geobacillus Stearothermophilus: Microorganism of choice for VHP.
  32. Ozone Sterilization

    • "O3"
    • FDA clearance in 2003.
    • No aeration needed.
    • Medical devices/surgical instruments can be used as soon as the 4 1/2 hour cycle is complete.
  33. Ozone Sterilization Process Parameters

    • Cycle Temperature: 87.4 F to 97 F
    • Ozone Concentration: 59%
    • Relative Humidity: 85% to 100%
    • Time: 4 hours and 30 mins.
  34. Safety

    • OSHA,FDA, and NIOSH regultae O3 as a toxic gas.
    • O3 = Bluish gas/pungent odor (odor threshold 0.003 to 0.01 parts per million/ppm)
    • O3 sterilizer design converts all O3 produced to oxygen before being exhausted into the room.
  35. Exposure Monitoring

    • O3 sterilization requires a well ventilated room with 10 air exchanges per hour.
  36. Materials Compatibility

    • O3 process is not recommended for sterilizing sealed ampules, liquids, natural rubber, latex and fabrics.
  37. Sterilizer Performance Monitors

    • Geobacillus Stearothermophilus: Microorganism of choice for O3.