  1. 1overview• What is mass spectrometry
    – Principles of ionisation and mass analysers
    • Interpretation
    – The mass spectrum
    – Interpretation – peaks and isotopes
    • Hyphenated methods GC-MS and LC-MS
    – Analysis in the pharmaceutical industry / forensics
  2. 2Principles of MSPrinciples of MS
  3. 3Examples mass spectraExamples mass spectra
  4. 4Interpretation: Mass, m/z, isotopes What is mass?• SI unit for mass is kg
    • Mass of a single carbon-12, 12C, atom is 1.99266×10-26 kg
    • One unit (u) of mass (atomic mass unit, amu) is defined as a Dalton
    (Da).
    • One Dalton is defined as 1/12 the mass of a single carbon-12 atom.
    • Thus, one 12C atom has a mass of 12.0000 Da
  5. 5Calculation:• Mass of a peptide is 5685 Da
    • m/z of double protonated peptide?
    • Mass of the peptide in kg?
  6. 6Calculation: • Mass of a peptide is 5685 Da • m/z of double protonated peptide?Calculation: • Mass of a peptide is 5685 Da • m/z of double protonated peptide?
  7. 7Calculation: • Mass of a peptide is 5685 Da • Mass of the peptide in kg?Calculation: • Mass of a peptide is 5685 Da • Mass of the peptide in kg?
  8. 8Resolution & mass accuracy on mellitinResolution & mass accuracy on mellitin
  9. 9Mass resolution– High Resolution Mass Spectrometry
    • If sufficient resolution (R > 5000) exists, mass numbers can be
    recorded to precise values (6 to 8 significant figures)
    – Resolution:
    • Ability to distinguish between ions of adjacent m/z values
    Mass resolution
  10. 10Mass resolution For example,Mass resolution For example,
  11. 11The Analysis of isotope patternsis helpful
    in element composition determination
  12. 12Interpretation: IsotopesMost elements have more than one stable isotope.
    For example, most carbon atoms have a mass of 12 Da, but in nature, 1.1%
    of C atoms have an extra neutron, making their mass 13 Da.
    Why do we care?
    Mass spectrometers can “see” isotope peaks if their resolution is high
    enough.
    If an MS instrument has resolution high enough to resolve these isotopes,
    better mass accuracy is achieved.
  13. 13Isotopes• Atoms of the same element can have different numbers of
    neutrons
    • The most common isotope of hydrogen has no neutrons at all;
    there's also a hydrogen isotope called deuterium, with one
    neutron, and another, tritium, with two neutrons.
    Isotopes
  14. 14Isotopes cont.Isotopes cont.
  15. 15Definitions for Monoisotopic ion mass Nominal ion mass Average ion mass• Monoisotopic ion mass
    Mass of ion for a given empirical formula calculated using the
    exact mass of the most abundant isotope of each element, as
    accurate as possible (e.g. 1H = 1.007825 and 12C=12.000000).
    • Nominal ion mass
    Mass of ion calculated using the integer mass (mass number
    of most abundant isotope of each element (e.g. 1H =1 and
    12C=12).
    • Average ion mass
    Mass calculated with the average of the isotopic mass of each
    element, weighted for isotopic abundance (e.g. H=1.00797
    and C=12.01115).
  16. 16Interpretation: isotopesInterpretation: isotopes
  17. 17Interpretation: isotopesInterpretation: isotopes
  18. 18Mass spectrum of peptide with multiple CatomsMass spectrum of peptide with multiple Catoms
  19. 19Mass spectrum of insulin (257 C-atoms)Mass spectrum of insulin (257 C-atoms)
  20. 20The nitrogen rule (use nominal mass)• If the molecular mass of an unknown compound to the
    nearest integer value is an odd number, the compound
    contains an odd number of nitrogens in its molecular formula.
    Correspondingly, if the molecular mass is an even number, the
    compound contains zero or an even number of nitrogens in its
    molecular formula
  21. 21The nitrogen rule (use nominal mass)The nitrogen rule (use nominal mass)
  22. 22Isotopes - summary• Mass spectra will show isotopic patterns of ions for the elements present
    • For n Carbon atoms, 13Cn will contribute to
    M+1 ion (n x 1.1%)
    • If one chlorine is present a 3:1 pattern will be
    seen for M and M+2
    • Hydrogen and oxygen have low abundances for their isotopes 2H and 18O and do not give significant signals for small molecules
  23. 23summaryMass Spectrometry can give structural evidence
    from:
    1. Isotope distribution and accurate mass
    leading to elemental composition
    2. Fragmentation; tandem mass spectrometry
    3. When combined with HPLC – retention time data
    4. Derivatization of functional groups
  24. 24Liquid Chromatography-Mass Spectrometry• Interfacing HPLC with mass spectrometerLiquid Chromatography-Mass Spectrometry
  25. 25Advances of LC-MS detection• A mass spectrometer is
    more sensitive and far
    more specific than all
    other LC detectors
    • Analysis of compounds
    without a suitable
    chromophore
    • Identification of
    components in
    unresolved
    chromatographic peaks
  26. 26Why Liquid Chromatography - Mass Spectrometry?• Two-dimensional abundance data
    and three-dimensional mass
    spectral data from a mass
    spectrometer
    • Mass spectral data add specificity
    that increases confidence in the
    results of both qualitative and
    quantitative analyses
    • Information gained: molecular
    weight, structure, identity, quantity,
    and purity of a sample
  27. 27Interfacing LC with MS Fundamental issues:• LC eluent – limited acceptable LC flow rate ~ 200
    uL/min (ESI)
    • Ionisation under high vacuum
    • Temperature differences
    • No non-volatile salts, e.g. phosphate, borate.
    • Ion competition (matrix effect)
    • LC is the time limiting step in high through put
    analysis
  28. 28Interfacing LC with MS contChanges to LC methods required for modern LC-MS
    systems generally involve changes in sample
    preparation and solution chemistry to:
    • Ensure adequate analyte concentration
    • Maximise ionisation through careful selection of
    solvents and buffers
    • Minimise the presence of compounds that compete
    for ionisation or suppress signal through gas-phase
    reactions
  29. 29Ionisation chemistry (ESI)Formation of analytes ions in solution is essential to
    achieving good ESI data
    • Select more volatile buffers to reduce the buildup of
    salts in the ion source
    • Adjust solvent pH according to the polarity of ions
    desired and the pH of the sample
    • Use solvents that have low heat of vaporisation and
    low surface tension to enhance ion desorption
    • Make sure that gas-phase reactions do not neutralise
    ions through proton transfer or ion pair reactions
  30. 30HPLC–ESI–MS/MS applications:• Now considered the benchmark for
    measurement of drugs and their metabolites
    in biological matrices
    • The method of choice for screening for
    inherited metabolic disorders
    • Drug quantification and therapeutic drug
    monitoring
    • Toxicological investigations
  31. 31Gas Chromatography-Mass Spectrometry• Identification and quantitation of (semi) volatile
    organic compounds in complex mixtures
    • Determination of MW/elemental composition of
    unknowns
    • Quantitation of drugs and their metabolites (blood,
    urine) for pharmacological and forensic applications
    • Environmental analysis – quantitation of pollutants in
    drinking and waste water
    • Explosive investigation
    • Steroids
    • ‘Specific tests’ in Forensic science
  32. 32Gas Chromatography-Mass SpectrometryGas Chromatography-Mass Spectrometry
  33. 33The column• Mobile phase: gas (helium or nitrogen)
    • Stationary phase: 5% phenyl-, 95% dimethylsiloxane
    • 25 m x 0.5 mm i.d. (capillary column of glass or
    stainless steel)
    • The column within the oven can be programmed to
    increase the temperature gradually (ramped up to ca.
    400°C by 1°C/min to 40°C/min)  better separation:
    as the temp. increases, low bp analytes elutes faster
    than high bp analytes  two separation forces:
    temperature and the stationary phase interactions
    • DETECTOR: MASS SPEP: Electron ionization (EI) and
    quadrupole mass analyze
  34. 34Sample preparation• Injection volume 0.5 - 2.0 µL
    • Analytical sensitivity of 1 – 100 pg per component
    • Derivatives
    In order to make the molecule both resistance to the high temperatures
    used in the GC and to selectively improve the mass spectra, functional
    groups can be ‘derivatised’ i.e. chemically modified. Reaction with a
    silylation reagent is most common
    Why Derivatize:
    • To permit analysis of compounds not directly amenable to analysis due to,
    for example, inadequate volatility or stability
    • Improve chromatographic behaviour or detectability
  35. 35The range of functional groups which can be derivatised is large and includes:COOH
    OH
    NH2
    C=C
    The essential aspect of the
    chemistry is that the
    reaction must be fast,
    efficient, and the reagents
    and by-products should be
    volatile and easily removed
  36. 36—