A2 Physics Medical Imaging

Sean Byrne Physics AS level English 61 cards

Diagram of production of X-rays

Cards in this deck (61)

  1. 1X-raysA form of EM radiation
    Produced when high-speed electrons hit metal targets
  2. 2Production of X-raysHeated filament undergoes thermionic emission releasing high speed electrons
    p.d. between cathode and anode causes electrons to accelerate
    Electrons bombard metal target emitting X-rays which leave through a windown
    Some kinetic energy of electrons transferred into metal target as thermal energy
    Metal target is cooled by water or spun around to increase target area
  3. 3Diagram of production of X-raysDiagram of production of X-rays
  4. 4Tube currentRate of arrival of electrons at metal target
  5. 5Intensity of X-raysDepends on no. of electrons hitting anode per unit time
    Proportional to the heater current
    A more intense X-ray produces an image quicker
  6. 6Hardness of X-raysDepends on acceleration of electrons
    Proportional to p.d. between anode and cathode
    A harder X-ray has more penetrating power
  7. 7How to reduce the radiation doseSoft X-rays cannot fully pass through patient, becomes useless but still contributes to the total radiation dose of the patient
    To reduce the radiation dose, cut off soft X-rays by placing an aluminum filer used to absorb them
  8. 8X-ray spectrumEmerge from the tube with a range of energies as represented in the spectra below
  9. 9Graph of X-ray spectrumGraph of X-ray spectrum
  10. 10Braking radiationX-ray photons released when electrons decelerate as it strikes anode, attracted by the nucleus of an atom in the anode and loses energy
  11. 11Characteristic radiationRearrangement of electrons in anode when a high-speed electrons strikes, excites orbital electrons which then de-excite, emitting photons and giving rise to spectrum lines; specific to material of anode
  12. 12X-ray imagingX-ray radiation plates in the same way as visible light
    Degree of blackening depends on total X-ray exposure
    Mostly used to distinguish bones from tissue because bones have a higher density than surrounding tissue
    Can be used to identify organs if the densities of surrounding tissue are sufficiently different
  13. 13SharpnessThe ease with which the edge of a structure can be determined
  14. 14How to increase sharpness in X-raysReduce the area of target anode
    Reduce aperture size: reduces beam width
    Place lead grid in front of photographic film: absorbs scattered X-rays and reduces partial image
  15. 15ContrastThe visual difference between the areas of blackening and light
  16. 16How to improve contrast of X-raysIncrease exposure time
    Use harder X-rays: increases penetration power
    Reduce scattering of X-ray beam
    Use fluorescent 'contrast medium'
  17. 17X-ray intensityIn medium where X-rays are absorbed, intensity of a parallel X-ray beam decreases by a constant fraction in passing through equal small thicknesses of the medium
    I = I(0)e^ux
    Where u is the linear absorption coefficient unique to medium
    x is the thickness of the medium
  18. 18Graph of X-ray intensity through a mediumGraph of X-ray intensity through a medium
  19. 19Half value thicknessThickness of the medium required to reduce the transmitted intensity to one half of its value
    Related to linear absorption coefficient by
    u = ln2 / xhalf
  20. 20Computed Tomography (CT scan)X-ray imaging only produces a two dimensional image with no impression of depth, cannot tell if tissue is near to the surface or deep within the body
    Tomography is a procedure which forms a 3 dimensional plane of the object
    Putting together several planes produces a whole 3 dimensional image which can be rotated
  21. 21Diagram of CT scanDiagram of CT scan
  22. 22VoxelA small cube in a three dimensional image
  23. 23The number in each square (voxel)The density that the computer will register for that section of the object
  24. 24How voxel in CT scans workAs the scanner goes around each part has a different density which the computer can model
    The computer then put these together to form a 3 dimensional shape
  25. 25Diagram of how voxels workDiagram of how voxels work
  26. 26How to create a well defined CT scanX-ray beams must be well collimated so that it consists of parallel rays - rays must no spread
    Detector must consist of regular array of tiny detecting elements - smaller the detector the better the image
  27. 27Advantages of CT scanProduce images that show three-dimensional relationships between different tissues
    Can distinguish tissues with quite similar densities
  28. 28UltrasoundAny sound wave that has a frequency above the upper limit of human hearing, 20kHz
  29. 29Piezo electric transducerCan be used to record and produce ultrasonic waves
  30. 30Diagram of a piezo-electric transducerDiagram of a piezo-electric transducer
  31. 31Generating ultrasonic wavesCharged atoms of a transducer in an electric field move close to oppositely charged plates and the overall crystal either compresses or extends
    When an alternating voltage with frequency f is applied to the crystal, it causes it to contact and expand at the same frequency of f
    This acts as the vibrating source of ultrasound waves
  32. 32Receiving ultrasonic wavesUltrasonic waves change pressure in medium
    Charged atoms in crystal shift position closer to plates
    Opposite charges induced in the silver plates
    Induced potential difference across the plates
    Potential difference fluctuates which can be processed
  33. 33Reflection in ultrasoundRequires ultrasonic waves to pass from one medium to another
    When a beam of ultrasound wave reaches a boundary between two different media, the beam is partially refracted and partially reflected
  34. 34Why I = Ir + ItConservation of energy
  35. 35Specific acoustic impedanceProduct of density of medium and speed of sound in medium
    Z = pc
    Between two media, the difference in acoustic impedance determines the fraction of incident intensity that is reflected
  36. 36Intensity reflection coefficientRatio of intensiyt of reflected wave and intensity of incident wave
    (Z2 - Z1)^2/(Z2 + Z1)^2
  37. 37Comparing acoustic impedanceVery large fraction reflected at air-tissue boundary
    Large fraction reflected at tissue-bone boundary
    Very little reflected at boundary between soft tissues
  38. 38Purpose of gel in ultrasonic scansVery little transmission at an air-skin boundary
    Almost complete transmission at a gel-skin boundary
    Because acoustic impedance of gel and skin are very similar
  39. 39Attenuation of ultrasonic wavesSimilar to X-rays, ultrasonic waves are also absorbed by the medium in which they are passing through
    I = I(0)e^-kx
    k is the absorption coefficient
  40. 40Ultrasound Imaging ProcedureTransducer is placed in contact with skin and a gel acting as a coupling medium
    Pulses of ultrasound are directed into the body
    The wave is reflected at boundary between tissues
    The reflected pulse is detected and processed
    The time for return of echo gives information on depth
    Amount of reflection gives information on structures
  41. 41A-scanMeasures distance of of different boundaries from transducer held in one position graphically
  42. 42B-scanRepeats A-scan from multiple angles to form 2 dimensional image
  43. 43How do atomic nuclei with an odd number of protons behave in a magnetic fieldLike tiny magnets
  44. 44What atmoic nuclei is used most in MRIHydrogen because it is abundant in all organic tissue
  45. 45When there is no magnetic field, protons are aligned:Randomly
  46. 46When a magnetic field is introduced, most protons...Most protons align themselves with 'north' facing 'south' - stable low energy state
    Some protons align themselves in the opposite way with 'south' facing 'south' - unstable high energy state
  47. 47Aligned protons are not stationary; they...spin on their axis of rotation
  48. 48Atomic spinA fundamental property of a subatomic particle that defines how the particle rotates on its axis
  49. 49PrecessionThe movement of the axis of rotation of a spinning object (proton) around another external axis
  50. 50Larmor frequencyThe angular frequency of the circular path of precession of the object (proton)
  51. 51Process of nuclear magnetic resonanceThe applied external magnetic field causes nuclei to precess
    Angular frequency of precession is called Larmor frequency
    Frequency of radio pulses applied equals to the Larmor frequency
    Causes hydrogen nuclei to resonate and 'flip' into higher energy state
    Larmor freq = radio freq = resonance freq
  52. 52Relaxtion of protonsThe protons in high energy state are unstable so must 'relax' and come back to lower energy state
    The excess energy is transmitted back as radio waves which can be detected
    The time taken for these radio waves to be detected determines the relaxation time
  53. 53Relaxation timeTime taken for a nucleus to fall back to a lower energy state
    Depends on environment of protons:
    water and watery tissues - several seconds - high
    Fatty tissues - hundreds of milliseconds - low
    Cancerous tissue - intermediate
  54. 54How MRI scanning worksA large uniform magnetic field causes all protons in body to have the same Larmor frequency
    A non-uniform magnetic field is applied to locate a particular position of a proton within the person
  55. 55Procedure of an MRIA strong constant magnetic field is applied along body
    Hydrogen nuclei precess about the direction of the field
    A radio frequency pulse is applied
    The pulse is at the Larmor frequency which causes resonance in hydrogen nuclei
    On relaxation, the nuclei de-excite and emit pulse of r.f.
    r.f. pulses are detected, processed and displayed
    A calibrated non-uniform magnetic field enables position of nuclei to be located and for location of detection to be changed
  56. 56Advantages of X-raySharp image
    Improvable contrast
    Can form image where air is trapped
  57. 57Drawbacks of X-rayEquipment heavy and not portable
    May cause ionization and damage tissues
  58. 58Benefits of ultrasoundPortable equipment
    Less harmful than X-ray
    Used for treatment of sprained joints
  59. 59Drawbacks of ultrasoundCannot be used to form image where air is trapped
    Image not sharp due to refraction
    Rapid movements in tissue - may damage
  60. 60Benefits of MRIClearer image than ultrasound
    Image can be studied in any plane or direction
    Lower health risk than X-ray/ultrasound
  61. 61Drawbacks of MRIEquipment heavy and not portable
    Requires patient to remain still
    If patient body contains metal, heated/attracted
    Pacemakers may be upset by magnetic field

Information

Description
AS level Physics Flashcards on A2 Physics Medical Imaging, created by Sean Byrne on 26/02/2018.