Arterial blood pressure
Sphygmomanometry is the
indirect measurement of blood pressure. To do it directly would involve
inserting a cannula attached to a diaphragm and transducer into the radial or
dorsalis pedis arteries. However, this is invasive. Sphygmomanometry uses the
brachial artery because it branches directly off the aorta and can be measured
at the level of the heart, so it is considered equal to aortic pressure. When
measuring the blood pressure, one hears:
·
The systolic
pressure, which is the highest pressure produced when the ventricles contract.
This is typically ~120 mmHg and heard as turbulent/noisy intermittent squirting
of blood as the high pressure of the cuff is released. It reflects the amount
of work the heart is being asked to perform with each beat to generate
pressure.
·
The diastolic
pressure, which is the lowest pressure produced when the ventricles fill. This
is typically ~80 mmHg and it is a continuous, smooth (laminar) flow of blood
only allowed when the cuff pressure is lower than diastolic pressure. It is
heard as the pulse sounds disappearing – previously the cuff had been inflated
above diastolic pressure and so the brachial artery was still closing and
opening and creating pulse sounds as blood squirts through. This reflects the
state of the blood vessels: both their stretchiness and how much resistance to
blood flow there is (to maintain blood pressure).
Auscultatory sphygmomanometry – relies on the sounds of blood through the artery.
Includes both using a stethoscope, and automatic blood pressure devices
containing a microphone, such as the DGS (PC programme in Phys lab which
produces graphs).
Oscillometric
sphygmomanometry – relies on shock
waves from the opening and closing of the artery changing the pressure in the inflated
cuff as they reach it. Is the more common technique in automatic blood pressure
monitoring; includes the Omron and Omega automated blood pressure monitors.
We first looked at blood pressure changes by attaching
a cuff around the brachial artery and analysing the pulse changes on the DGS.
The way the cuff pressure oscillates as the cuff is deflated can be used to
estimate systolic and diastolic pressures.
We then measured blood pressure manually (using an
Aneroid cuff and stethoscope kit). I was unable to find my partner’s pulse, or
mine.
Next, we used the Omega oscillometric automatic blood
pressure monitor. We compared this with an Omron oscillometric automatic blood
pressure monitor, but were unable to compare either of the automated methods
with the manual.
Class data:
·
Using a paired T-test, there was no
significant difference between the manual and automated, and between the
different automated, methods of measuring blood pressure – EXCEPT in the
diastolic manual measurement vs. both the Omron and Omega measurement. One
explanation could be slow reaction to sounds disappearing in the manual method.
·
Using an unpaired T-test, there was strong
evidence of a difference between male and female systolic pressure, evidence of
a difference between male and female diastolic pressure, and no evidence of a
difference between male and female heart rates.
·
Using a paired T-test, there was no evidence
of a difference in systolic pressure, diastolic pressure, or heart rate when
the class held their breath while these measurements were made, than when
breathing was normal. This could have been a result of a lack of controlled
experimental conditions.
Venous pressure
Venous pressure is very
low (only a few mmHg) because of dissipation of ventricular pressure energy
throughout the vascular tree from pressure to heat energy, and the large
diameter of the vessels providing little resistance. The normal range is 0-6 mmHg.
We measured it by holding our arms at our sides (so the veins in the backs of
our hands would fill with blood) then raising and lowering our arms to find the
height at which the vessels collapse and refill. We then measured this vertical
distance (in mm) from the level of the hand to the heart (4th rib),
and converted these mm into mmHg (mm x 12.9, as Hg is 12.9x heavier than
blood).
Class data:
Not available.
My results were inaccurate (an average of
7.75 mmHg) as I did not mark the two heights and work out the difference, or
ask somebody to help me measure, but rather estimated the difference using a
ruler.
Pathologies
High arterial/venous
blood pressure: high intrathoracic pressure, arterial stenosis, hypervolaemia,
obstruction, heart failure. Low arterial/venous blood pressure: hypovolaemia,
low intrathoracic pressure.