CH 07: Decontamination: Point-of-Use Preparation and Transport
Slides in this set (6)
-
Goals of Point-of-Use Prep & Transport
- Removal of gross soil.
- Prevention of damage.
- Prevention of cross contamination.
- Keeping others safe.
-
- Instrument preparation begins at the point of use.
-
Point-of-Use Preparation Guidelines
- Remove gross soil.
- Follow the manufacturer's instructions for point-of-use cleaning and preparation.
- Separate reusable sharps from other instruments.
- Separate reusable linen.
- Remove disposable components.
- Open hinged instruments, disassemble multi-part instruments and place them in trays. Heavy: bottom, Light: top, Cords, Scopes, Cameras: Very top.
- Keep items together. (Sets, multi-part items)
-
- Keep instruments moist and prevent soil from drying on their surface. Use foams, gels, sprays, or a moist towel. Limit the time kept in a soak basin or solution to prevent instrument damage.
- Empty fluids from contaminated device containers using facility protocols.
- Remove, contain, or separate third-party reprocessing items and send to CS, per facility protocols.
- For reusable instruments used in a case with CJD notify CS and Infection Prevention Control. Follow hospital procedures.
- Notify CS: Items needing repair. Tag them to remove them from the system.
- Notify CS: Items requiring turnover/turnaround.
-
- Transport soiled instruments in enclosed carts. (Preferred)
- Open Carts: Must have a solid bottom/ cover cart during transport.
- Don't place metal trays/containers on top of plastic trays/containers.
- Don't use containers/carts transporting contaminated items to transport clean and sterile items. (Unless the cart is decontaminated between uses)
-
- AAMI recomends practices for transporting contaminated items.
- Procedures should also reflect OSHA regulations.
- AORN provides additional information.