CH 15: Low-Temperature Sterilization
Slides in this set (37)
-
- Ethylene Oxide (EtO)
- Hydroden Peroxide (H2O2)
- Ozone (O3)
-
Low Temperature Sterilization Requirements
- Efficacy (Effectiveness): Provides the minimum-required Sterility Assurance Level (SAL).
- Safety: No toxic residues on packaging/devices upon completion of sterilization cycles.
- Exposure Monitoring.
- Sterilization Performance Monitoring.
- Penetration.
- Material Compatibility.
- Adaptability.
- Approval.
-
- FDA regulated.
- FDA/CDC outlines the SAL for each sterilization method.
-
Toxicity Standards: Low-Temp Sterilants
- PEL: Permissible Exposure Limits; parts per million (ppm). By OSHA.
- TWA: Time Weighted Average (over an 8 hour day). By OSHA.
- IDLH: Immediately Dangerous to Life or Health. National Institute for Occupational Safety and Health (NIOSH) established.
- EtO: OSHA PEL/TWA (1.0 ppm), NIOSH IDLH limit (800 ppm). Monitoring required.
- H2O2: OSHA PEL/TWA (1.0 ppm), NIOSH IDLH limit (75 ppm). Monitoring is not required.
- O3: OSHA PEL/TWA (0.1 ppm), NIOSH IDLH limit (5 ppm). Monitoring is not required.
-
- Eto inactivates microbes (alkylation).
- H2O2/O3 destroys microbes (oxidation).
-
- For heat and moisture-sensitive medical devices.
- High vapor pressure and a low boiling point of 51.3 F
- EtO cartridges are placed inside chambers of less than 10 cubic feet. The cartridge is punctured under vacuum.
- Sterilization Cycle 5 Stages: Preconditioning and humidification, gas introduction, exposure, evacuation and air washes.
- Cycle: About 2 1/2 hours, excluding aeration time.
- After sterilization items must go through an aeration process (removes residual EtO before the chamber can be unlocked for item removal).
-
- OSHA classifies EtO as a carcinogen and reproductive hazard.
- EtO Sterilizers: Located in a well ventilated area/ air exchange rate of at least 10 changes per hour/ in negative-pressure rooms.
-
EtO Sterilization Process Parameters
- Gas Concentration: 45 to 1,200 mg/l
- Temperature: 98 F to 145 F
- Relative Humidity: 40% to 80%
- Exposure Time: 1 to 6 hours
- Aeration Time: 8 to 12 hours at 122 F to 140 F
-
- Devices affixed to the employee's clothing/breathing zone (within 1 foot of the person's nose) tests airborne EtO concentrations.
-
- EtO should not be used to sterilize: 1) Liquids, 2) Devices with energy sources (sparks are created), 3) Leather Items (chemicals used in the tanning process will combine with EtO and form chlorohydrin)
-
- EtO paper/plastic or Tyvek peel pouches, approved fabric wrappers, medical crepe paper, polypropylene and most container systems.
- Never use aluminium foil, cellophane, nylon films, polyester, PVC (plastic) films and styrofoam.
-
- Vacuum pulse remove EtO.
- Warm air circulates to remove residuals from instruments. Minimum 8 hours at 140 F and 12 hours at 122 F.
-
Sterilizer Performance Monitors
- Bacillus atrophaeus: Microorganism of choice for EtO.
-
- Hydrogen Peroxide Gas Plasma: Oxidative Process. Safer than EtO, rapid cycle times. Byproducts of the cycle (water and oxygen) are nontoxic = no aeration phase.
- Gas plasma are highly ionized gases.
-
- H2O2 Gas Plasma: Hydrogen Peroxide solution from 59% to 95% is proven effective at killing microorganisms.
-
- H2O2 gas plasma sterilizers sterilize some single channel flexible endoscopes, cameras, rigid endoscopes, light cords, batteries and power drills.
-
Types of H2O2 Gas Plasma Systems
- Compact Systems: 28-38 min processing times.
- Large Capacity Systems: 75 min cycle times.
-
Sterilization Cycle/Process Parameters
- Phases of hydrogen peroxide gas plasma.
- 1)Vacuum: The load is heated. Removes remaining water as it evaporates.
- 2)Injection: H2O2 into chamber.
- 3)Diffusion: Drives H2O2 vapor into small crevices/lumens of devices.
- 4)Plasma: A vacuum decreases the pressure/radio frequency (RF) energy is radiated from an electrode screen. RF energy ionizes the H2O2 creating H2O2 gas plasma.
- 5)Vent: No aeration. Use instruments after the 24-75 min sterilization cycle.
-
H2O2 Gas Plasma:Process Parameters
- Time: 24-75 mins.
- Cycle Temperature: Less than 131 F
- Hydrogen Peroxide Concentration: 59% to 95%
- Vacuum Level: Automatically controlled.
- Plasma: Automatically controlled.
-
- H2O2 liquid irritates skin/damages eyes.
- In vapor phase: Irritates eyes,nose, throat, and lungs.
- H2O2 is packaged in sealed cassettes. Chemical leak indicators on them change from yellow or white to red.
- After 5 sterilization cycles the cassette is automatically ejected into a collection box for safe disposal.
- Latex, vinyl (PVC), or nitrile gloves are worn to remove items from a cancelled cycle.
-
- H2O2 gas plasma is not compatible with:
- 1) Liquids and powders.
- 2) Any material that absorbs liquid.
- 3) Items that contain cellulose, such as cotton, paper or cardboard, linens, huck towels, gauze sponges or any item containing wood pulp.
-
- Tvek pouches are compatible.
- Don't use cellulose-containing packaging such as paper/plastic pouches.
-
Sterilizer Performance Monitors
- Geobacillus Stearothermophilus: Microorganism of choice for H2O2 gas plasma.
- AORN Guidelines: Daily biological monitoring.
-
- "VHP": used since 2007.
- 59% hydrogen peroxide solution.
- Injected 4 times during each sterilization cycle. After the 4th injection period, the load is automatically aerated in the sterilizer. VHP is converted to water and oxygen.
-
Types of Vaporized H2O2 Systems
- Single Preprogrammed 55 Min Cycle: Lumened and nonlumened instruments and devices.
- Two Preprogrammed Cycles: 28 min (non lumened instruments) and 55 min (instruments with lumens/non stainless steel mated surfaces)
-
Sterilization Cycle Process Parameters
- Phases of VHP systems include:
- 1) Conditioning: Removes air/excess moisture then recharges with dry sterile air.
- 2) Leak Test: Vacuum is held to ensure a leak tight chamber.
- 3) Sterilization: 4 injection pulses, each followed by a hold period.
- 4) Aeration: Automatic.
-
Vaporized H2O2: Process Parameters
- Cycle Tempeerature: Less than 122 F
- H2O2 Concentration: 59%
- Time: 28 to 55 mins.
-
- Concentrated H2O2 is corrosive to skin, eyese, nose, throat, lungs, and gastrointestinal tract.
- Wear latex, vinyl (PVC), or nitrile gloves when removing items from a cancelled load.
-
- Vaporized H2O2 sterilization is compatible with telescopes, cameras, light cables. batteries and many lumened devices.
- The VHP system is not intended to process liquids, linens, powders, or any cellulose materials.
-
- Polywrap, nonwoven sterilization packaging made of 100% polypropylene, Tyvek.
-
- Geobacillus Stearothermophilus: Microorganism of choice for VHP.
-
- "O3"
- FDA clearance in 2003.
- No aeration needed.
- Medical devices/surgical instruments can be used as soon as the 4 1/2 hour cycle is complete.
-
Ozone Sterilization Process Parameters
- Cycle Temperature: 87.4 F to 97 F
- Ozone Concentration: 59%
- Relative Humidity: 85% to 100%
- Time: 4 hours and 30 mins.
-
- OSHA,FDA, and NIOSH regultae O3 as a toxic gas.
- O3 = Bluish gas/pungent odor (odor threshold 0.003 to 0.01 parts per million/ppm)
- O3 sterilizer design converts all O3 produced to oxygen before being exhausted into the room.
-
- O3 sterilization requires a well ventilated room with 10 air exchanges per hour.
-
- O3 process is not recommended for sterilizing sealed ampules, liquids, natural rubber, latex and fabrics.
-
Sterilizer Performance Monitors
- Geobacillus Stearothermophilus: Microorganism of choice for O3.