Test
Training exam
Crea una cuenta gratis para terminar este test
Guarda tu progreso y consulta tu resultado final.
¿Ya tienes cuenta? Inicia sesión
Preguntas del test (9)
1What are the requirements for the DICOM Purge feature in order to work?
- AImages need to be cleared form the local PIC folder.
- BImages need to be archived to a PACS
- CImages need to be cleared form the local PIC folder OR Images need to be archived to a PACS
- DImages need to be cleared form the local PIC folder AND Images need to be archived to a PACS
2What is required to enable the QC in Eye Care PACS
- A1. HL7 orders must be supported and activated.
2. User need to belong to SymphonyWebAutors Group - B1. HL7 must be supported and activated.
2. User need to belong to SymphonyWebQCUsers Group - C1. HL7 orders must be supported and activated.
2. User need to belong to SymphonyWebQCUsers Group
- A1. HL7 orders must be supported and activated.
3What is needed to configure the OISDFC.INI for DICOM archive of the DCZ container files
- ASOP Class and Transfer Syntax
- BSOP Class and Study UID
- CInstance UID & Transfer Syntax
- DInstance UID & Study UID
4What type of HL7 messages do we support?
- AADT, ORU, MDM, QRY, ORM, QBP, ACK
- BORM, ORU, ADT, QRY, MFN, ACK
- CBAR, ADT, ORU, MFN, ACK, QRY, QBP
5Message Trigger Events for HL7.
What are the activities that should cause the Information System (IS) to initiate a message:- AThe IS interface sends an order message (ORM) to EyeStation, when an order is created,
updated, or cancelled. This message usually contains patient and physician information along with
requested observation details. - BThe IS interface sends an ADT message to EyeStation, whenever a patient's demographics are
updated, or a patient is merged, or to send a response to demographics query. - CThe IS interface sends an MFN message to EyeStation, whenever a physician information is
added/updated.
- AThe IS interface sends an order message (ORM) to EyeStation, when an order is created,
6Message Trigger Events
The following activities cause the EyeStation to initiate a message:
The EyeStation sends an order result message (ORU) to IS, whenever an order is fulfilled.
● The EyeStation sends an MDM or ORU message to IS, when a digital report is created or updated.
● The EyeStation sends a query message (QRY) to IS, whenever patient validation and additional
demographic data is required.
● The EyeStation sends an ACK message (positive or negative acknowledgement) to IS, when an
ORM, ADT, or MFN message is received and processed.- AVerdadero
- BFalso
7HL7 Message Rule file. What is the meaning of the different types of brackets?
- A{ } Curly braces indicate that a segment can repeat
[ ] Square brackets indicate that a segment is optional - B{ } Curly braces indicate that a segment is optional
[ ] Square brackets indicate that a segment can repeat
- A{ } Curly braces indicate that a segment can repeat
8With SegDef file we define what segments we expect
- How many
- What order
- Which messages do we use to add/update or merge.- AVerdadero
- BFalso
9What are the HASP bits needed for a MECP server with 8 users, import form IOLMaster, Spectralis, Nidek Micro-Perimeter, Oculus Pentacam, RetCam
- AWinstation Version Number 11, SymphonyWeb enabled, Symphony Web Number of Licenses: 8, Montage enabled, Auto History, Auto Stabilizer, Perfect View , Net HASP, OCT, VFA, ULT, MO1, CTP
- BWinstation Version Number 11, SymphonyWeb enabled, Symphony Web Number of Licenses: 8, Review, Montage enabled, Auto History, Auto Stabilizer, Perfect View, Net HASP OCT, VFA, ULT, MO1, CTP
- CWinstation Version Number 11, SymphonyWeb enabled, Symphony Web Number of Licenses: 8, Montage enabled, Auto History, Auto Stabilizer, Perfect View OCT, VFA, ULT, MO1, CTP
Información
- Descripción
- Examination after the training
- Autor
- Filip Carpentier
- Idioma
- English
- Visitas
- 197
- Etiquetas
- Valoración
- Aún sin valoraciones Valora este test