Nucear APPs 2 - Radiation detection
Diapositivas del conjunto (14)
-
Gas detectors; Ionisation detector

- Incoming particles leave ionisation track
- voltage across 2 plates; electrons to anode, positive ions to cathode
- small current generated is proportional to the rate of energy loss in gas
-
-

- Similar to gas but the anode is a thin wire
- For typical values additional ionisation occurs resulting in electron ionising everything it hits
- almost all ionisation occurs near wire; pulse shape independent of where energy was dumped
- fast collection time (10ms) and large amplification possible
- output pulse is proportional to input energy over broad region of initial voltage
-

- similar to proportional counter but operates at higher voltage
- De-excitation releases uv photons which ionise other atoms, extending the region of ionisation along wire.
- produces large discharge which eventually stops with buildup of positive charge around the wire
- +ve ions drift to cathode where high prob of electron being released, retriggering the counter.
- counter this with a quench gas (eg methane, ethane) which exchanges charge with +ve ions and disassociates harmlessly
-
general features of gas detectors
- rugged
- cheap
- gas is low density so inefficient for gamma and fast neutrons
- efficient for charged
- efficient for thermal neutrons if 3-He or isotopic enriched 10-BF3 used as fill gas
- coat inner surface with 235-U ; thermal neutron fissions the 235-U, fission product recoil into gas
- fast neutrons use a moderator polythene ~10cm diameter surrounds and reduces neutron to thermal energy (bonner sphere)
- Geiger counters have low max count rate ~2000 cps and have finite lifetime due to quench gas used
-

- Radiation deposits energy in scintillator (usually gamma or xray) creating excitation region
- The de-excitation releases many photons; 1 Mev gamma ¬ 5*10^4 photons, E= 20ev. Reflection shield helps capture light on photo cathode
- pulse of current at anode is proportional to # of photons produced which is proportional to energy of incoming radiation (gamma or xray)
- resoultion deltaE/E ~ 7%
-
- Organic: transparent to its own light due to band gaps.
- There is a rapid production of flurorescent light
- Inorganic: Pure crystals have filled valence band ( large band gap - insulator). Not transparent to its own radiation so impurity atoms ( activators) are added. Tl and NaI to provide intermediate states in band gap. Crystals are transparent to light from activator states
-
- Inorganic: for gamma rays need high density of -e, ie high atomic number and high material density.
- NaI- sodium iodide
- BGO- Bismuth germanate
- LaBr3- Lanthanum bromide, good resolution
- Not great resolution (NaI ~6-10%), better for LaBr3
- 1st interaction for 1Mev gamma will be elastic, if leaves detector is compton spectrum edge
- can also backscatter
- or be fully absorbed - full energy photopeak

-
- Organic: good for B-rays and fast neutrons but too low Z and density for gamma rays ( due to much lower probability interaction -comp scattering)
- chance of detecting the scattered photon in same crystal or liquid is also low . few events in full E peak
- Good points for scintillators:
- high gain, large signal output
- cheap
- Bad points:
- fragile
- cracks in fast temp changes
- gain drifts; temp and magnetic affects PMT
-
- accurate - used in research
- is a single crystal - Si or Ge or CdZnTe
- voltage applied between electrodes ( p type and n-type)
- like a solid ( not gas ) ionisation chamber. Radiation creates electron hole pairs which are collected by the contacts. charge collected is proportional to E of radiation
- small E gap - cooled to liquid N2 ( 77K ) to minimise thermal excitations
- good resolution - deltaE/E ~ 1/ (N) ^ 1/2
-
- very pure crystal with thin p and n type crystals at either end
- Si- Low Z
- good for charged particles and x rays
- DealtaE ~ 20kev for 5Mev alpha
- finite lifetime due to damage from heavy ions
- need cooling: but cooling improves resolution
- Ge - high Z
- used for gamma rays
- DealtaE ~ 2 kev at 1.33Mev
- liquid N2 required for cooling
- high purity, large volume crystal - expensive
- 'CZT' - CdZnTe
- used for gamma and xray
- good a room temp
- worse resolution than Ge currently

-
- p type ( acceptor) and n type (donor) are in contact.
- Electrons from n type diffuse into p type until E field produced stops flow
- small depletion region is created ( no free charge available)
- this region is the detector, can increase it by applying an additional reverse bias.
- If radiation is deposited in the depleted region, free charge is created which is immediately swept away to the contacts by the E field.
- charge pulse that is measured is proportional to the gamma ( or whatever radiation) Energy
- can also see extra peaks in NaI detector due to e-e+ annihilation: E + 511 kev if one gamma escapes + 2*511 kev if both are detected