  1. Isothermal Titration Calorimetry (ITC)

    • Differential Scanning Calorimetry (DSC)
      • finds the denaturation temperature and how much energy it takes to unfold a molecule
        • Proteins unfolding is usually cooperative, meaning when it goes it all goes at once
          • When its only one peak and its sharp, cooperative
      • Temperature is gradually increased (~1℃/min) by supplying electrical energy to heaters, in such a way so Tref=Tsample
        • What is recorded: the difference in energy required to heat the sample to keep the same temperature as the reference
      • can be used to compare thermal stability of different proteins
    • Used for investigating thermodynamics of ligand binding
    • Experimental arrangement:
      • 1) Ligand is added to sample a few microliters at a time, at ~1 minute intervals (titrations) → releases heat
        • 2) Electrical energy is used to keep sample and reference at same temperature (re-equilibrate the sample)
    • The ITC trace is used to find: n, ^H, and Keq (and Kd)
      • ^G can be calculated from Keq
        • S can be calculated using ^H and ^G
    • Gives you all thermodynamic parameters
    • *A very small amount of heat is released in each ligand binding event, so a large amount of sample must be used so there is enough heat released to be detectable*
  2. Surface Plasmon Resonance (SPR)

    • What happens to light at the interface of 2 transparent materials with different index of refraction: it gets refracted
      • At low angle of incidence, there is “total internal reflection”
        • There is an angle of incidence where the energy of the refracted ray can be directed along the interface.
          • At the SPR angle, energy can be transferred to electrons in the gold film, creating an excited group of electrons called a “plasmon”
            • If the angle of incident light is adjusted just right so that the refracted ray is almost parallel to the gold film, the gold atoms in the film will absorb energy from the refracted light ray and some of the electrons in the gold atoms will become excited.
              • When the excited electrons in the gold atoms become mobile it's called a plasmon
            • At the SPR angle, there is a decrease in the intensity of the reflected light, due to the transfer of energy to the plasmon
            • The SPR angle depends on the index of refraction at the gold surface, which depends on the protein and ligand concentration at the interface
    • Detector measures the intensity of the reflected light
      • If protein #1 binds to protein #2, there will be an “SPR response”
        • “SPR response” means a change in the SPR angle, described in response units (RU)
    • For SPR to be useful in measuring protein-protein (or protein-ligand) interactions, it is necessary to attach a target molecule to the gold surface
      • Histidine-tagged proteins can be attached to the surface using their ability to bind to metals with 2+ charge
    • Procedure:
      • 1. Analyte region contains buffer only. Protein #1 is attached to surface
        • 2. Introduce protein #2 into solution
          • 3. Binding kinetics → kon
            • 4. Wash away protein #2
              • 5. Dissociation kinetics → koff
                • ratio of Koff to Kon is Kd
    • Results:
      • SPR data obtained at a range of temperatures can be used to find H and S for ligand binding, using a van’t Hoff analysis
        • SPR can measure much smaller values of Kd than ITC
          • SPR also uses much less material than ITC
  3. Differential Scanning Fluorescence (DSF)

    • Use a dye (SYPRO) that increases its fluorescence in the presence of unfolded protein
      • Ligand binding often stabilizes protein structure
        • An increase in denaturation temperature indicates that a protein has bound a ligand
    • A way to detect the denaturation temperature of proteins
    • qualitatively detects ligand binding
    • typically used in quickly screening a large number of ligands for binding to a protein
  4. Equilibrium Dialysis

    • Porous membrane lets ligand pass through, and keeps M inside
    • Assumptions in equilibrium dialysis:
      • All binding sites have equal affinity for ligands
      • Binding of ligands is reversible
    • kd can be calculated from the concentration of [L] and [M] on either side of the membrane
  5. Fluorescence Anisotropy

    • Measuring Fluorescence Anisotropy
      • Measure the fluorescence emission with a polarizing filter orientated parallel (and then perpendicular) to the polarization direction of the excitation light
        • Anisotropy = 0 when fluorescent light is not polarized
        • Anisotropy 0 when fluorescent light is at least partially polarized
        • The fluorescence emission of a ligand becomes more polarized when bound to a protein
    • Fluorescence anisotropy is better at detecting binding of a small fluorescent ligand to a large protein
    • If a fluorophore does NOT rotate during its fluorescent lifetime, the emitted light will be polarized
    • If the fluorophore rotates during its fluorescent lifetime, the emitted light will be mostly depolarized

Imágenes y archivos del mapa

  • Methods for Measuring Ligand Binding

Información

Descripción
Biology Mapa Mental sobre Methods for Measuring Ligand Binding, creado por Eva Stettler el 12/05/2022.
Creado por
Eva Stettler
Visitas
0
Valoración

Escríbenos

¿Sobre qué es tu mensaje?

¿Qué te parece el nuevo diseño?

¿Te gusta?