Principles of IRR
Inhalt der Mindmap
Principles of IRR
1. Registration with health & safety executive
2. Radiation risk assessment
3. Local rules
- Must be written
- Must contain contingencies for a radiological accident
- Must name the employer, RPA and RPS
- Must say the controlled area
- 1.5m of x-ray tube and patient in all directions
4. Appointing RPS
- Usually a GDP in practice
- Ensures compliance of IRR
5. Appointing RPA
- Appointed in writing
- Must be a medical physicist
- Checks installation of equipment
- Decides whether to report wrong exposures to CQC
6. Monitoring of areas and people
- ALARP
- When to dose monitor staff
- When taking over 100 intra-orals or 50 DPTs per week
- Who to notify if there is an overexposure
- Patient
- RPA
- Doctor
7. Equipment installation
- New intra-oral sets should be 60-70kV
- 2mm switch cable
- Rectangular collimation
- Physics QA intervals at maximum every 3 years
- Critical examination performed by installer and acceptance test